FDA Adverse Event Death Summary report: N

MOSS MIAMI

MDR report key: 6907298 · Received October 2, 2017

Report

Report Number
1526439-2017-10779
Event Type
Death
Date Received
October 2, 2017
Date of Event
August 31, 2017
Report Date
September 6, 2017
Manufacturer
DEPUY SYNTHES SPINE
Product Code
HWR
UDI-DI
10705034194455
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A COMPLAINT INVESTIGATION WILL BE PERFORMED. THE COMPLAINT PRODUCT IS NOT AVAILABLE FOR THE INVESTIGATION. A SUPPLEMENTAL REPORT IS NOT ANTICIPATED UNLESS THE RESULTS OF THE COMPLAINT INVESTIGATION IDENTIFY A CORRECTIVE ACTION OR ADDITIONAL RELEVANT INFORMATION. SHOULD THE PRODUCT BECOME AVAILABLE, A PHYSICAL EVALUATION WILL BE CONDUCTED AND A SUPPLEMENTAL REPORT FILED WITH THE RESULTS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEVICE NOT RETURNED FOR EVALUATION.

Description of Event or Problem · 1

SURGEON INFORMED ME ABOUT THE VIPER SURGERY HE PERFORMED ON (B)(6) 2017. THE FEMALE PATIENT HAD BEEN PROVIDED WITH 5 VIPER SCREWS ON EACH SIDE. SINCE ONLY 8 EXTENSIONS WERE AVAILABLE IN THE SET, THE SURGERY WAS PERFORMED IN A KIND OF MINI OPEN TECHNIQUE. ONE SCREW ON EACH SIDE WAS PLACED WITHOUT AN EXTENSION. BESIDES 2 EXTENSION DISASSEMBLED DURING THE SURGERY AND 2 SCREWS EACH SIDE WERE PLACED WITHOUT EXTENSIONS. WHEN THE ROD PLACING WAS FINISHED (ROD WELL PRE-BENDED), THE INNIES WERE PLACED. THE ROD COULDN'T BE PLACED FAR ENOUGH INTO THE SCREW HEAD AND THE INNIE COULDN'T BE TIGHTENED. THEREFORE PLUNGER (CODE NO. 275015900) MM ROD INTRODUCTION INSTRUMENT WAS USED. SURGEON TRIED TO PLACE THE ROD WITH THE PLUNGER INTO THE SCREW HEAD IN ORDER TO BE ABLE TO TIGHTEN THE INNIE. IN DOING THIS THE SURGEON SLIPPED OFF THE ROD AND THE PATIENT'S AORTA WAS INJURED. THE PATIENT DIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
686348 MOSS MIAMI DRIVER, PROSTHESIS HWR DEPUY SYNTHES SPINE 275015900 10705034194455

Patients

Seq Age Sex Outcome Treatment
1 Death