PARADYM

MDR report key: 6905272 · Received September 29, 2017

Report

Report Number
1000165971-2017-00728
Event Type
Injury
Date Received
September 29, 2017
Date of Event
August 31, 2017
Report Date
September 1, 2017
Manufacturer
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
Product Code
MRM
PMA / PMN Number
P980049
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PLEASE REFER TO THE ATTACHED ANALYSIS REPORT.

Description of Event or Problem · 1

REPORTEDLY, ON (B)(6) 2017, EARLY BATTERY DEPLETION WAS REPORTED FOR THE SUBJECT DEVICE. THE BATTERY VOLTAGE AND MAGNET RATE HAD DECREASED FROM 2.92V AND 87 MIN-1 ON (B)(6) 2017 TO 2.68V AND 82 MIN-1 ON (B)(6) 2017. THE FOLLOWING WARNING MESSAGES WERE DISPLAYED: '[27] THE DEVICE WAS REINITIALIZED 2 TIMES SINCE THE BEGINNING OF LIFE. LAST RESET DATE: (B)(6) 2017 4:41, [18] R.R.T (RECOMMENDED REPLACEMENT TIME) DETECTED: PLAN TO REPLACE DEVICE." THE BATTERY CURVE WAS REPORTEDLY SHOWING THAT THE BATTERY VOLTAGE ONCE REACHED THE END OF SERVICE LEVEL. BATTERY REFORMING WAS PERFORMED ON (B)(6) 2017. REPORTEDLY, THE PATIENT DID NOT UNDERGO TO ANY EXAMINATIONS SUCH AS MRI SINCE THE LAST FOLLOW-UP. THE DEVICE WILL BE REPLACED ON (B)(6) 2017. PRELIMINARY ANALYSIS CONFIRMED THAT THE ICD REACHED ITS RRT POINT. PATIENT CARE RECOMMENDATIONS WERE PROVIDED ACCORDINGLY TO REPLACE THE ICD. IT WAS ALSO CONFIRMED THAT TWO RESETS OCCURRED ON 15 AUG 2017 AND TOTAL CHARGE TIME SINCE IMPLANTATION WAS AROUND 3 MINUTES. THE DEVICE WAS EXPLANTED ON 08 SEPTEMBER 2017 AND WILL BE RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

REPORTEDLY, ON (B)(6) 2017, EARLY BATTERY DEPLETION WAS REPORTED FOR THE SUBJECT DEVICE. THE BATTERY VOLTAGE AND MAGNET RATE HAD DECREASED FROM 2.92V AND 87 MIN-1 ON (B)(6) 2017 TO 2.68V AND 82 MIN-1 ON (B)(6) 2017. THE FOLLOWING WARNING MESSAGES WERE DISPLAYED: "[27] THE DEVICE WAS REINITIALIZED 2 TIMES SINCE THE BEGINNING OF LIFE. LAST RESET DATE: (B)(6) 2017 4:41 [18] R.R.T (RECOMMENDED REPLACEMENT TIME) DETECTED: PLAN TO REPLACE DEVICE." THE BATTERY CURVE WAS REPORTEDLY SHOWING THAT THE BATTERY VOLTAGE ONCE REACHED THE END OF SERVICE LEVEL. BATTERY REFORMING WAS PERFORMED ON (B)(6) 2017. REPORTEDLY, THE PATIENT DID NOT UNDERGO TO ANY EXAMINATIONS SUCH AS MRI SINCE THE LAST FOLLOW-UP. THE DEVICE WILL BE REPLACED ON (B)(6) 2017. PRELIMINARY ANALYSIS CONFIRMED THAT THE ICD REACHED ITS RRT POINT.PATIENT CARE RECOMMENDATIONS WERE PROVIDED ACCORDINGLY TO REPLACE THE ICD. IT WAS ALSO CONFIRMED THAT TWO RESETS OCCURRED ON (B)(6) 2017 AND TOTAL CHARGE TIME SINCE IMPLANTATION WAS AROUND 3MINUTES. THE DEVICE WAS EXPLANTED ON (B)(6) 2017 AND WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
685102 PARADYM DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER MRM SORIN GROUP ITALIA S.R.L. - CRM FACILITY PARADYM DR 8550 2549

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention