FDA Adverse Event Injury Summary report: N

VERSA-DIAL/COMP TI STD TAPER

MDR report key: 6904738 · Received September 29, 2017

Report

Report Number
0001825034-2017-07592
Event Type
Injury
Date Received
September 29, 2017
Date of Event
August 25, 2017
Report Date
December 28, 2017
Manufacturer
ZIMMER BIOMET, INC.
Product Code
MBF
PMA / PMN Number
PK060716
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 118001, VERSA-DIAL/COMP TI STD TAPER, 297720. 115330, BIOMOD/COMP REV BASE STD, 436740. 115310, COMP RVRS SHLDR GLNSP STD 36MM, 593670. 11-113706, BIO-MOD ST.10X115 W/ALIGN HOLE, 936090. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 - 2017 - 07589. 0001825034 - 2017 - 07590. 0001825034 - 2017 - 07591.

Additional Manufacturer Narrative · 1

UPON REASSESSMENT OF THE REPORTED EVENT, IT WAS DETERMINED THAT THIS DEVICE WAS NOT INVOLVED IN THE EVENT. THE INITIAL REPORT WAS SUBMITTED IN ERROR AND SHOULD BE VOIDED.

Description of Event or Problem · 1

IT WAS REPORTED A PATIENT UNDERWENT SHOULDER REVISION APPROXIMATELY SIX YEARS POST-IMPLANTATION DUE TO PAIN, DYSFUNCTION, ASEPTIC LOOSENING, ELEVATED METAL ION LEVELS, AND ADVERSE LOCAL TISSUE REACTION. DURING THE REVISION, THE GLENOSPHERE AND BASEPLATE WERE FOUND TO BE LOOSE. EXTENSIVE WEAR ON TRUNNION WAS NOTED, WHICH IS SUSPECTED TO BE THE SOURCE OF THE METALLOSIS AND METAL ION ELEVATION. ALL COMPONENTS WERE REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
684620 VERSA-DIAL/COMP TI STD TAPER PROSTHESIS, SHOULDER MBF ZIMMER BIOMET, INC. N/A 297700

Patients

Seq Age Sex Outcome Treatment
1 63 YR Hospitalization| R