FDA Adverse Event
Malfunction
Summary report: N
CSK-TC ELECTRODE
MDR report key: 6901675
·
Received September 28, 2017
Report
- Report Number
- 3006630150-2017-03829
- Event Type
- Malfunction
- Date Received
- September 28, 2017
- Date of Event
- September 5, 2017
- Report Date
- September 5, 2017
- Manufacturer
- COSMAN MEDICAL
- Product Code
- GXI
- UDI-DI
- 00813250010114
- PMA / PMN Number
- K050084
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
EXPIRATION DATE: NA. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED: MODEL #: CSK-TC10 LOT #: 041217. DESCRIPTION: CSK TC ELECTRODE, 10CM TOTAL QUANTITY: 1.
Description of Event or Problem · 1
DEVICE ANALYSIS OF THE RETURNED ELECTRODES INDICATED THAT TRACES OF POTTING EPOXY WERE FOUND ON THE HUBS. THE EPOXY LEAKED THROUGH THE SHAFT OPENINGS IN IT'S LIQUID STATE DURING THE MANUFACTURING PROCESS, AND WAS NOT REMOVED AFTER CURING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682078 | CSK-TC ELECTRODE | ELECTRODE | GXI | COSMAN MEDICAL | CSK-TC15 | 080315 | 00813250010114 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |