FDA Adverse Event Malfunction Summary report: N

CSK-TC ELECTRODE

MDR report key: 6901675 · Received September 28, 2017

Report

Report Number
3006630150-2017-03829
Event Type
Malfunction
Date Received
September 28, 2017
Date of Event
September 5, 2017
Report Date
September 5, 2017
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250010114
PMA / PMN Number
K050084
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EXPIRATION DATE: NA. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED: MODEL #: CSK-TC10 LOT #: 041217. DESCRIPTION: CSK TC ELECTRODE, 10CM TOTAL QUANTITY: 1.

Description of Event or Problem · 1

DEVICE ANALYSIS OF THE RETURNED ELECTRODES INDICATED THAT TRACES OF POTTING EPOXY WERE FOUND ON THE HUBS. THE EPOXY LEAKED THROUGH THE SHAFT OPENINGS IN IT'S LIQUID STATE DURING THE MANUFACTURING PROCESS, AND WAS NOT REMOVED AFTER CURING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682078 CSK-TC ELECTRODE ELECTRODE GXI COSMAN MEDICAL CSK-TC15 080315 00813250010114

Patients

Seq Age Sex Outcome Treatment
1