FDA Adverse Event Injury Summary report: N

VIDAS® HS TROPONIN I

MDR report key: 6900246 · Received September 28, 2017

Report

Report Number
3002769706-2017-00286
Event Type
Injury
Date Received
September 28, 2017
Date of Event
September 2, 2017
Report Date
October 30, 2017
Manufacturer
BIOMERIEUX, SA
Product Code
MMI
PMA / PMN Number
K063243
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF REPEATABLE FALSE POSITIVE RESULTS IN ASSOCIATION WITH VIDAS® HIGH SENSITIVE (HS) TROPONIN (TN) I (REF. 415386) - LOT 1005783570 / 180720-0. AFTER TWO FALSELY ELEVATED HS TROPONIN I RESULTS, A (B)(6) YEAR-OLD MAN WAS SENT TO THE EMERGENCY ROOM. THE INITIAL SAMPLE WAS LATER CONFIRMED NEGATIVE WITH THE ROCHE TECHNIQUE. AN INTERNAL BIOMÉRIEUX INVESTIGATION WAS PERFORMED. RECORDS ANALYSIS: THE PRODUCT QUALITY LABORATORY HAS STUDIED THE CONTROL CARDS OF 16 BATCHES OF VIDAS® TNHS KIT : LOT 1005783570 / 180720-0 IS IN THE TREND ANALYSIS COMPARED TO THE OTHER BATCHES. THE QUALITY PRODUCT LABORATORY HAS TESTED 5 INTERNAL SAMPLES : ALL THE RESULTS WERE WITHIN THEIR SPECIFICATION. SO THE VIDAS® TNHS LOT 1005783570 / 180720-0 IS WITHIN ITS ACCEPTANCE CRITERIA. THE CUSTOMER'S SAMPLE WAS FORWARDED TO THE R&D DEPARTMENT FOR FURTHER INVESTIGATIONS. THE SAMPLE WAS TESTED ON : VIDAS TNI REF. 30448. VIDAS TNHS REF. 415386, BATCH 180602-0 (BATCH NUMBER DIFFERENT THAN THE ONE MENTIONED BY THE CUSTOMER). WITH VIDAS® TNI REF. 30448, THE RESULT WAS POSITIVE 0.81¿G/L. WITH VIDAS® TNHS REF. 415386, THE RESULT WAS 1478.0 NG/L. A DILUTION TEST WAS PERFORMED IN ORDER TO CHECK IF THE POTENTIAL INTERFERENCE COULD BE ELIMINATED. THE RESULTS OBTAINED AFTER DILUTION INCREASED COMPARED TO THE RESULT OBTAINED ON THE UNDILUTED SERUM. THIS IS IN FAVOR OF AN INTERFERENCE, BUT THIS BEHAVIOR WAS DIFFERENT FROM THAT USUALLY ENCOUNTERED. AN HBT TEST (HETEROPHILIC BLOCKING TUBE) WAS PERFORMED. THE RESULTS IN TNIU AND TNHS WERE POSITIVE. THIS IS NOT AN INTERFERENCE DUE TO HETEROPHILIC ANTIBODIES. UNFORTUNATELY, THE LACK OF VOLUME DID NOT ALLOW US TO IDENTIFY THE NATURE OF THIS INTERFERENCE. HOWEVER AS THE VIDAS® TNHS BATCH NUMBER USED BY QUALITY PRODUCT LABORATORY IS DIFFERENT THAN THE CUSTOMER'S BATCH, WE CAN EXCLUDE AN INTERFERENCE LINKED TO THE BATCH. VIDAS® TNHS LOT 1005783570 / 180720-0 PERFORMED WITHIN ITS ACCEPTANCE CRITERIA. THE PATIENT SAMPLE HAS INTERFERENCE WITH THE VIDAS® TNHS PARAMETER. IN THE INSTRUCTIONS FOR USE - LIMITATIONS OF THE METHOD SECTION, IT'S WRITTEN THAT "INTERFERENCE MAY BE ENCOUNTERED WITH CERTAIN SAMPLES CONTAINING ANTIBODIES DIRECTED AGAINST REAGENT COMPONENTS. FOR THIS REASON, ASSAY RESULTS SHOULD BE INTERPRETED TAKING INTO CONSIDERATION THE PATIENT'S CLINICAL HISTORY, AND THE RESULTS OF ANY OTHER TESTS PERFORMED."

Description of Event or Problem · 1

A CUSTOMER IN (B)(4) NOTIFIED BIOMÉRIEUX OF FALSE POSITIVE RESULTS IN ASSOCIATION WITH VIDAS® HIGH SENSITIVE TROPONIN I (REF. 415386) - LOT 1005783570. THOUGH THE VIDAS® HIGH SENSITIVE TROPONIN I ASSAY (REF. 415386) IS NOT REGISTERED IN THE UNITED STATES, A SIMILAR PRODUCT (VIDAS® TROPONIN I ULTRA ASSAY, REF. 30448) IS REGISTERED. THE PATIENT IS A (B)(6) MAN WITH CHEST PAIN. HIS PHYSICIAN HAD ORDERED THE HIGH SENSITIVE TROPONIN I TESTING, WHICH WAS COLLECTED AND TESTED ON (B)(6) 2017. THE FIRST RESULT WAS POSITIVE AT 1274.4NG/L. IT WAS RETESTED, AND FOUND POSITIVE AGAIN AT 1247.1NG/L. THE RESULT WAS REPORTED TO THE PHYSICIAN, AND THE PATIENT WAS SENT TO THE HOSPITAL EMERGENCY ROOM. THERE, HE HAD ANOTHER TROPONIN TEST (THE TECHNIQUE USED IS NOT KNOWN), WHICH WAS NEGATIVE. THE PATIENT WENT HOME AS A RESULT OF THE NEGATIVE RESULT OBTAINED IN THE HOSPITAL. PER THE CUSTOMER, HE RECEIVED NO TREATMENT AND HAD NO CLINICAL SIGNS OF INFARCTION. ON (B)(6) 2017, THE CUSTOMER SENT THE PATIENT SAMPLES DRAWN ON (B)(6) 2017 TO ANOTHER LABORATORY. THE RESULTS HAVE BEEN CONFIRMED NEGATIVE WITH THE ROCHE TECHNIQUE. A BIOMÉRIEUX INTERNAL INVESTIGATION HAS BEEN INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680821 VIDAS® HS TROPONIN I VIDAS® HS TROPONIN I MMI BIOMERIEUX, SA 1005783570

Patients

Seq Age Sex Outcome Treatment
1 39 YR Hospitalization