3M CUROS DISINFECTING CAP FOR NEEDLELESS CONNECTORS
Report
- Report Number
- 2110898-2017-00128
- Event Type
- Malfunction
- Date Received
- September 26, 2017
- Date of Event
- August 17, 2017
- Report Date
- September 26, 2017
- Manufacturer
- 3M HEALTH CARE
- Product Code
- LKB
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
CUSTOMER REPORTED THE CATALOG NUMBER FOR THIS REPORT WAS CFF1-270 (OR CFF 10-250). CFF1-270 ARE INDIVIDUAL CAPS AND CFF10-250 ARE STRIPS(10 CAPS /STRIP). OPERATOR OF DEVICE WAS UNKNOWN. CUSTOMER REPORTED THE PATIENT'S HICKMAN CATHETER WAS USED FOR DAILY HOME INFUSIONS OF TPN. TYPE OF REPORTABLE EVENT. THIS REPORT DID NOT INVOLVE A DEATH, SERIOUS INJURY OR A MALFUNCTION. MALFUNCTION WAS SELECTED FOR THIS REPORT BECAUSE A SELECTION WAS REQUIRED FOR THIS SECTION OF THE REPORT. THE CUSTOMER REPORT NOTED THAT CUROS CAPS ARE NOT INTENDED TO ATTACH TO A FEMALE LUER HUB AND ARE DESIGNED TO BE USED ON NEEDLELESS CONNECTORS. THE PRODUCT PACKAGING INCLUDES THE FOLLOWING WARNING AND CAUTIONARY STATEMENTS: WARNING: TO AVOID POTENTIAL INJURY - USE ONLY ON NEEDLELESS CONNECTORS. THE PACKAGE INSERT STATES THE FOLLOWING INFORMATION RELATED TO INTENDED USE AND INSTRUCTIONS FOR USE: INTENDED USE: THE CUROS¿ DISINFECTING CAP IS INTENDED FOR USE ON NEEDLELESS CONNECTORS ONLY AS A DISINFECTING CLEANER PRIOR TO I.V. ACCESS AND TO ACT AS A COVER BETWEEN LINE ACCESSES...... INSTRUCTIONS FOR USE: WARNING: TO AVOID POTENTIAL FOR INJURY - USE ONLY ON NEEDLELESS CONNECTORS...... IN ADDITION, 3M PROVIDES TRAINING MATERIALS (INCLUDING GRAPHICS) INSTRUCTING CUSTOMERS TO APPLY THE CUROS CAP ONLY TO NEEDLELESS CONNECTORS AND NOT TO APPLY THE CUROS CAP DIRECTLY TO A CATHETER HUB. IN SUMMARY, THE PRODUCT PACKAGING AND INSTRUCTIONS FOR USE CLEARLY STATE VIA A WARNING THAT THE PRODUCT SHOULD ONLY BE USED ON NEEDLELESS CONNECTORS.
AN INTERVENTIONAL RADIOLOGIST REPORTED AN OUTPATIENT PRESENTED WITH A NONFUNCTIONING HICKMAN CATHETER LINE. A SPONGE FROM A CUROS CAP WAS FOUND IN THE FEMALE LUER CATHETER HUB AND WAS REMOVED. NO INJURY WAS REPORTED, THE PATIENT WAS BEING MONITORED FOR SIGNS AND SYMPTOMS OF INFECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 674363 | 3M CUROS DISINFECTING CAP FOR NEEDLELESS CONNECTORS | CUROS DISINFECTING CAP FOR NEEDLELESS CONNECTORS | LKB | 3M HEALTH CARE | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |