FDA Adverse Event Malfunction Summary report: N

23 G X 1 IN. BD ECLIPSE¿ 3 ML BD LUER-LOK¿ SYRINGE WITH DETACHABLE NEEDLE

MDR report key: 6892120 · Received September 25, 2017

Report

Report Number
8041187-2017-00118
Event Type
Malfunction
Date Received
September 25, 2017
Date of Event
July 31, 2017
Report Date
October 24, 2017
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FMI
PMA / PMN Number
K161170
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: ONE PICTURE RECEIVED AND ONE ACTUAL SAMPLE WITHOUT PACKAGING WAS RETURNED FOR INVESTIGATION. OBSERVED THE SYRINGE BARREL WAS DAMAGED. DHR REVIEW: NO ABNORMALITY AND QN REPORTED DURING ECLIPSE ASSEMBLY PRODUCTION. REVIEWED SYRINGE DHR FOR CATALOGUE# S20351SGT, BATCH# 7142288, 7087209, 7108087, 7142041, 7115368. NO QUALITY NOTIFICATION WAS RAISED FOR ALL THE BATCHES ON THE REPORTED NONCONFORMANCE. NO SIMILAR NONCONFORMANCE WAS DETECTED IN IN-PROCESS AND OUTGOING INSPECTION. INVESTIGATION CONCLUSION: ONE PICTURE AND ONE ACTUAL SAMPLE WITHOUT PACKAGING RECEIVED. WE OBSERVED DAMAGE ON THE SYRINGE BARREL. ROOT CAUSE DESCRIPTION: PROBABLE CAUSE COULD BE DUE TO POOR THROW OUT OF THE ASSEMBLED SYRINGE AT THE AUTO-REJECT STATION. THE BARREL COULD HAVE BEEN SLANTED AND HIT THE SIDE GUIDE RESULTING IN DAMAGE BARREL. CAPA# (B)(4) HAD BEEN INITIATED TO ADDRESS DAMAGED SYRINGE NONCONFORMANCE.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT AS REPORTED THAT THE BARREL ON A 23 G X 1 IN. BD ECLIPSE¿ 3 ML BD LUER-LOK¿ SYRINGE WITH DETACHABLE NEEDLE WAS CRACKED WHICH COULD LEAD TO LEAKAGE. NO INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
670890 23 G X 1 IN. BD ECLIPSE¿ 3 ML BD LUER-LOK¿ SYRINGE WITH DETACHABLE NEEDLE SAFETY ENGINEERED SYRINGE AND NEEDLE FMI BECTON DICKINSON MEDICAL (SINGAPORE) 7135159

Patients

Seq Age Sex Outcome Treatment
1 Other