1823260-2017-02033
Report
- Report Number
- 1823260-2017-02033
- Event Type
- Malfunction
- Date Received
- September 21, 2017
- Date of Event
- August 30, 2017
- Report Date
- September 21, 2017
- PMA / PMN Number
- K060373
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
UNIQUE IDENTIFIER (UDI)#: (B)(4).
THE CUSTOMER RECEIVED A QUESTIONABLE ELECSYS HCG + BETA TEST SYSTEM RESULT FOR ONE PATIENT SAMPLE. THE INITIAL RESULT WAS 0.100 MIU/ML WITH A DATA FLAG AND WAS REPORTED OUTSIDE THE LABORATORY TO THE DOCTOR WHO QUESTIONED THE RESULT. THE REPEAT RESULT WITH A 1:100 DILUTION WAS 20539 MIU/ML. THE SAMPLE WAS ALSO REPEATED ON ANOTHER ANALYZER WITH A 1:100 DILUTION AND THE RESULT WAS 21049 MIU/ML. THE REPEAT RESULTS WERE BELIEVED TO BE CORRECT. THERE WAS NO ADVERSE EVENT. THE REAGENT LOT NUMBER WAS 210151. THE EXPIRATION DATE WAS REQUESTED BUT WAS NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE COULD NOT FIND A CAUSE. HE DID FIND THE SIPPER NOZZLE SPRING MISSING FROM THE ASSEMBLY AND THE SAMPLE PROBE ASSEMBLY SET SCREWS WERE LOOSE WHICH CAUSED A SMALL AMOUNT OF PLAY IN THE HORIZONTAL MOTION. HE TIGHTENED THE SET SCREWS, REPLACED THE MISSING SPRING, AND REALIGNED THE SAMPLE PROBE ASSEMBLY. HE RAN PERFORMANCE TESTING AND THE RESULTS WERE WITHIN SPECIFICATIONS. THE CUSTOMER RAN CALIBRATION AND QC WHICH WERE ALL WITHIN SPECIFICATIONS. AS THE CUSTOMER DID NOT REPORT ANY FURTHER ISSUES, IT WAS DETERMINED THE ISSUE WAS RESOLVED BY THE SERVICE ACTIONS.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |