HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 6881944 · Received September 21, 2017

Report

Report Number
3004123209-2017-01087
Event Type
Malfunction
Date Received
September 21, 2017
Date of Event
September 11, 2017
Report Date
October 4, 2017
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K041067
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER).

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). ROUTINE TESTING CONFIRMED THAT THE PAD-PAK WAS FIRST WAS INSTALLED BY THE CUSTOMER ON (B)(6) 2009. THE DEVICE RECORDS MANUAL POWER UPS OF TEN MINUTES IN DURATION BETWEEN (B)(6) 2017. INVESTIGATION FOUND THE FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 0

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE SWITCHES ON AUTOMATICALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
665119 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD SAM N/A

Patients

Seq Age Sex Outcome Treatment
1