FDA Adverse Event Injury Summary report: N

LITTLE PAL STIRRUP

MDR report key: 6877362 · Received September 19, 2017

Report

Report Number
1221538-2017-00004
Event Type
Injury
Date Received
September 19, 2017
Date of Event
January 4, 2017
Report Date
August 18, 2017
Manufacturer
ALLEN MEDICAL SYSTEMS, INC.
Product Code
FWZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ROOT CAUSE IS EXPECTED WEAR OF THE PADS AT 5 YEARS PAST USEFUL LIFE. THE STIRRUP FUNCTIONED PROPERLY AND PASSED ALL TESTING. THE PADS WERE FLATTENED AND VISIBLY DEVOID OF CUSHIONING AND SHOULD NOT HAVE BEEN IN USE PER INSTRUCTIONS FOR USE " WARNING: TO PREVENT PATIENT AND/OR USER INJURY AND/OR EQUIPMENT DAMAGE, EXAMINE THE DEVICE FOR POTENTIAL DAMAGE OR WEAR PRIOR TO USE. DO NOT USE THE DEVICE: IF DAMAGE IS VISIBLE, IF PARTS ARE MISSING, OR IF IT DOES NOT FUNCTION AS EXPECTED."

Additional Manufacturer Narrative · 1

AWAITING DEVICE RETURN FOR INVESTIGATION OF ANY POSSIBLE CONTRIBUTION TO INJURY. A FOLLOW UP REPORT WILL BE FILED.

Description of Event or Problem · 1

THE CASE WAS PERFORMED IN (B)(6) 2017 ON A (B)(6) YEAR OLD MALE PATIENT USING THE LITTLE PAL STIRRUPS. PATIENT WAS IN FOR AN SWENSON PULL THROUGH AND OSTOMY CLOSURE AND THE CASE LASTED 7 HRS. PATIENT DEVELOPED COMPARTMENT SYNDROME OF LEFT LEG REQUIRING FASCIOTOMIES OF LEFT CALF, DEEP AN SUPERFICIAL COMPARTMENTS WITH APPLICATION OF A WOUND VAC. HE HAD SUPERFICIAL CLOSURES OF FASCIOTOMIES 3 DAYS LATER WITH DEBRIDEMENT AND IRRIGATION. THE PATIENT HAS MADE A FULL RECOVERY. THEY RECALL THAT THE INJURY SEEMED TO BE WHERE THE EDGE OF THE STIRRUP WOULD MEET THE CALF.

Description of Event or Problem · 1

THE CASE WAS PERFORMED IN (B)(6) 2017 ON A (B)(6) MALE PATIENT USING THE LITTLE PAL STIRRUPS. PATIENT WAS IN FOR AN SWENSON PULL THROUGH AND OSTOMY CLOSURE AND THE CASE LASTED 7 HOURS. PATIENT DEVELOPED COMPARTMENT SYNDROME OF LEFT LEG REQUIRING FASCIOTOMIES OF LEFT CALF, DEEP AN SUPERFICIAL COMPARTMENTS WITH APPLICATION OF A WOUND VAC. HE HAD SUPERFICIAL CLOSURES OF FASCIOTOMIES 3 DAYS LATER WITH DEBRIDEMENT AND IRRIGATION. THE PATIENT HAS MADE A FULL RECOVERY. THEY RECALL THAT THE INJURY SEEMED TO BE WHERE THE EDGE OF THE STIRRUP WOULD MEET THE CALF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
657555 LITTLE PAL STIRRUP STIRRUP FWZ ALLEN MEDICAL SYSTEMS, INC. A-10057-A1

Patients

Seq Age Sex Outcome Treatment
1 6 YR Required Intervention