FDA Adverse Event Other Summary report: N

GENERAL ELECTRIC MEDICAL SYSTEMS

MDR report key: 68721 · Received January 10, 1997

Report

Report Number
68721
Event Type
Other
Date Received
January 10, 1997
Date of Event
July 25, 1996
Report Date
August 6, 1996
Manufacturer
GE MEDICAL SYSTEM
Product Code
IZL
Report Source
User Facility report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

RADIOLOGY TECHNICIAN RECEIVED ELECTRICAL SHOCK WHILE PLUGGING IN PORTABLE X-RAY MACHINE. STATED THAT THERE WAS A SPARK AND LOUD "POP" SOUND. EMPLOYEE RECEIVED CARE IN THE ER. INVESTIGATION REVEALED WIRING BROKEN INTERNAL TO PLUG AND EVIDENCE OF A SHORT. PLUG WAS CUT OFF AND REPLACED. UNIT WAS RETURNED TO SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENERAL ELECTRIC MEDICAL SYSTEMS PORTABLE X-RAY MACHINE IZL GE MEDICAL SYSTEM AMX 110 *

Patients

Seq Age Sex Outcome Treatment
1 30 YR Other