FDA Adverse Event
Other
Summary report: N
GENERAL ELECTRIC MEDICAL SYSTEMS
MDR report key: 68721
·
Received January 10, 1997
Report
- Report Number
- 68721
- Event Type
- Other
- Date Received
- January 10, 1997
- Date of Event
- July 25, 1996
- Report Date
- August 6, 1996
- Manufacturer
- GE MEDICAL SYSTEM
- Product Code
- IZL
- Report Source
- User Facility report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
RADIOLOGY TECHNICIAN RECEIVED ELECTRICAL SHOCK WHILE PLUGGING IN PORTABLE X-RAY MACHINE. STATED THAT THERE WAS A SPARK AND LOUD "POP" SOUND. EMPLOYEE RECEIVED CARE IN THE ER. INVESTIGATION REVEALED WIRING BROKEN INTERNAL TO PLUG AND EVIDENCE OF A SHORT. PLUG WAS CUT OFF AND REPLACED. UNIT WAS RETURNED TO SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENERAL ELECTRIC MEDICAL SYSTEMS | PORTABLE X-RAY MACHINE | IZL | GE MEDICAL SYSTEM | AMX 110 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR | Other |