FDA Adverse Event Malfunction Summary report: N

640G INSULIN PUMP MMT-1711B

MDR report key: 6871234 · Received September 15, 2017

Report

Report Number
2032227-2017-48916
Event Type
Malfunction
Date Received
September 15, 2017
Date of Event
August 31, 2017
Report Date
February 3, 2020
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169662544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE RECEIVED WITH BLANK DISPLAY DUE TO CORRODED ELECTRONIC ASSEMBLIES. UNABLE TO PERFORM DISPLACEMENT TEST, SELF TEST, AND CURRENT MEASUREMENTS DUE TO DISPLAY ANOMALY.

Additional Manufacturer Narrative · 1

(B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2017 16:23:30 PST ICE BATCH USER (BATCH_ICE) (B)(6) COMPLAINT STATUS: IN PROCESS MDT INITIAL CONTACT: (B)(6) TAKEN BY: HULTER2 FATHER OF PT CALLED DURING OOH. PT DIT CHANGE THA BASAL SETTINGS, WHEN HE CHECKS THE PUMP AFTER 30 MINUTES THE PUMP HAS A BLANK SCREEN. PT CHANGED THE THE BATTERY, SCREEN STILL BLANK. PUMP HAS NOG BUMPED OR DROPPED. NO RADIATION OF MRI SCAN. CHANGED PUMP BY TNT. (B)(6). TH INPUT DATE: (B)(6) 2017 WARRANTY START: 07/21/2015 WARRANTY END: 07/20/2019 BATCH - NG1098723H.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
648059 640G INSULIN PUMP MMT-1711B PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1711B HG12CN8 00643169662544

Patients

Seq Age Sex Outcome Treatment
1