FDA Adverse Event Death Summary report: N

WELCH - ALLYN MRL

MDR report key: 686566 · Received November 26, 2005

Report

Report Number
686566
Event Type
Death
Date Received
November 26, 2005
Date of Event
November 11, 2005
Report Date
November 11, 2005
Manufacturer
MEDICAL RESEARCH LABS
Product Code
LDD
Report Source
User Facility report
Reporter Location
NM, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

TRANSPORTING A MALE WITH MI TO THE ER. VIFIB CODE 5 MINUTES OUT FROM HOSP. CONNECTED TO DEFIB PADS/DEFIBRILLATOR. MRL DEFIBRLLLATOR DID NOT DELIVER CHOSEN JOULES TO PT. DEFIBRILLATOR INDICATED CORRECT DESIRED JOULES TO BE DELIVERED. ACTUAL DELIVERED JOULES NOTED ON MRL LOG PRINTOUT NOTED NO SHOCKS 1 JOULE DELIVERED TO PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WELCH - ALLYN MRL MRL PC50 LDD MEDICAL RESEARCH LABS PIC 50 *

Patients

Seq Age Sex Outcome Treatment
1 52 YR Death