FDA Adverse Event Malfunction Summary report: N

1628664-2017-00339

MDR report key: 6865171 · Received September 14, 2017

Report

Report Number
1628664-2017-00339
Event Type
Malfunction
Date Received
September 14, 2017
Report Date
October 23, 2017
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVE TROPONIN-I RESULTS ON TWO PATIENTS. WHEN THE SAMPLES WERE RE-TESTED ON A DIFFERENT ANALYZER THE RESULTS WERE NEGATIVE. DURING A SUBSEQUENT SITE VISIT THE ABBOTT FIELD SERVICE ENGINEER (FSE) OBSERVED THAT THE MANIFOLD KIT VALVE (ROHS) (PART NUMBER 7-77612-03) AND THE 2-WAY BYPASS VALVE (ROHS) (PART NUMBER 7-200607-01) DID NOT WORK PROPERLY. THE FSE REPLACED THESE PARTS WHICH RESOLVED THE ISSUE. A REVIEW OF THE ARCHITECT SYSTEM OPERATIONS MANUAL ADDRESSES OPERATIONAL PRECAUTIONS AND LIMITATIONS, AS WELL AS, INFORMATION REGARDING TROUBLESHOOTING ERRATIC/DISCREPANT RESULTS. A REVIEW OF THE ARCHITECT I1SR02087 SERVICE HISTORY IDENTIFIED NO CONTRIBUTING FACTORS ON OR AROUND THE DATE OF THE COMPLAINT. THERE HAVE BEEN NO SUBSEQUENT CONTACTS FROM THE CUSTOMER REGARDING ERRATIC OR DISCREPANT RESULTS SINCE THE PARTS WERE REPLACED. A REVIEW OF TICKETS FOR THE MANIFOLD KIT VALVE (ROHS) (PART NUMBER 7-77612-03) OR THE 2-WAY BYPASS VALVE (ROHS) (PART NUMBER 7-200607-01) IDENTIFIED NO ADDITIONAL COMPLAINTS OR TRENDS FOR THE ISSUE. BASED ON THE INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE ARCHITECT SN I1SR02087 NOR THE MANIFOLD KIT VALVE (ROHS) OR THE 2-WAY BYPASS VALVE (ROHS).

Description of Event or Problem · 1

THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT TROPONIN RESULTS ON TWO PATIENTS. THERE WERE NO ACTUAL VALUES PROVIDED. THE CUSTOMER REPORTED THE ARCHITECT HS TROPONIN GENERATED POSITIVE RESULTS AND WHEN THE SAMPLES WERE RETESTED ON ANOTHER ANALYZER THE RESULTS WERE NEGATIVE. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED.

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT STAT HIGH SENSITIVE TROPONIN-I,| LIST # 03P25-26, LOT # UNKNOWN