FDA Adverse Event Malfunction Summary report: N

NITINOL TC ELECTRODE

MDR report key: 6863597 · Received September 13, 2017

Report

Report Number
3006630150-2017-03584
Event Type
Malfunction
Date Received
September 13, 2017
Date of Event
September 5, 2017
Report Date
September 5, 2017
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250013054
PMA / PMN Number
K050084
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EXPIRATION DATE: NA MODEL: TCN-10 LOT: 102716 ¿ QUANTITY ONE LOT: 102416 ¿ QUANTITY ONE LOT: 31517 ¿ QUANTITY TWO MODEL DESCRIPTION: NITINOL TC ELECTRODE, 100 MM TOTAL QUANTITY OF ELECTRODES - FOUR.

Description of Event or Problem · 1

THE RETURNED ELECTRODES WERE ANALYZED AND VISUAL INSPECTION ON ALL FOUR ELECTRODES FOUND THAT THE EPOXY HAS CHIP OUT AND DISCOLORATION. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642464 NITINOL TC ELECTRODE ELECTRODES GXI COSMAN MEDICAL TCN-10 102716 00813250013054

Patients

Seq Age Sex Outcome Treatment
1