BD 20 ML SYRINGE WITH 18G X 1.5 IN. NEEDLE
Report
- Report Number
- 3002682307-2017-00040
- Event Type
- Malfunction
- Date Received
- September 12, 2017
- Date of Event
- August 18, 2017
- Report Date
- October 3, 2017
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
INVESTIGATION: ROOT CAUSE ANALYSIS: WE CONCLUDE THAT THE CAUSE OF THE PROBLEM WAS PRODUCED AS A CONSEQUENCE OF A DAMAGE IN THE PLUNGER LIP. THIS COULD BE PRODUCED DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR IN THE PLUNGER ASSEMBLY MACHINE. CONFIRMATION: THE PROVIDED SAMPLE PRESENTED THE REPORTED ISSUE. WE COULD CONFIRM THE REPORTED ISSUE. CAPA DETERMINATION: NO - BASED ON AN EVALUATION OF SEVERITY AND OCCURRENCE IT WAS DETERMINATE THAT NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. SAMPLE EVALUATION: WE HAVE BEEN PROVIDED WITH THE AFFECTED SAMPLE. A LEAKAGE THROUGH THE PLUNGER ROD WAS OBSERVED IN THIS PROVIDED SAMPLE. AFTER THAT WE COULD DETERMINATED A DAMAGED IN THE PLUNGER ROD BY THE EVALUATION OF THE PLUNGER WITH MAGNIFICATION. BHR REVIEW: WE HAVE REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QA NOR NCMR'S. SYRINGES WERE PACKED IN MACHINE Nº2011 (SEPTEMBER 14 - 15TH, 2016). SYRINGES WERE ASSEMBLED IN MACHINE, Nº4252, AND Nº4251, IN LOT #6249011 (SEPTEMBER 7 - 17TH, 2016). RESEARCH HAS FOUND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE. WE HAVE ALSO REVIEWED THE BARREL LOTS #6249468, AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. WE HAVE ALSO REVIEWED THE PLUNGER LOTS #6253447, #6249472, AND #6239284 AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND.
IT WAS REPORTED THAT THERE WAS LEAKAGE FROM THE PLUNGER OF THE BD 20 ML SYRINGE WITH 18G X 1.5 IN. NEEDLE, DURING USE. NO REPORTED MEDICAL INTERVENTION OR SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 637506 | BD 20 ML SYRINGE WITH 18G X 1.5 IN. NEEDLE | 301948 | FMF | BECTON DICKINSON, S.A. | 1609154 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |