EVIS EXERA II COLONOVIDEOSCOPE
Report
- Report Number
- 8010047-2017-01329
- Event Type
- Malfunction
- Date Received
- September 12, 2017
- Date of Event
- August 11, 2017
- Report Date
- September 12, 2017
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FDF
- PMA / PMN Number
- PK100584
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. BUT WAS RETURNED TO OLYMPUS (B)(4). FOLLOWING ADDITIONAL HIGH LEVEL DISINFECTION AT (B)(4), THE SUBJECT DEVICE WAS SENT TO A THIRD PARTY LABORATORY FOR ADDITIONAL MICROBIOLOGICAL TESTING. IN THE ADDITIONAL TEST, THE TESTING INDICATED NO MICROBIAL GROWTH FOR THE SUBJECT DEVICE. OMSC REVIEWED THE MANUFACTURING HISTORY OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING ROUTINE SURVEILLANCE CULTURING TEST BY THE FACILITY, THE SUBJECT DEVICE TESTED POSITIVE FOR BACTERIA. THE CHANNELS OF THE SUBJECT DEVICE EXCEPT FOR THE AUXILIARY CHANNEL TESTED POSITIVE FOR FOLLOWING CFU OF BACTERIA. (B)(6). THE CUSTOMER REPORTED THAT THE SUBJECT DEVICE WAS REPROCESSED ACCORDING TO THE INSTRUCTION FOR USE AND THE FRENCH REGULATION. THE SUBJECT DEVICE HAD BEEN DISINFECTED USING AN OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR MODEL ETD-4 (NOT AVAILABLE IN THE U.S.) WITH PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 637343 | EVIS EXERA II COLONOVIDEOSCOPE | COLONOVIDEOSCOPE | FDF | OLYMPUS MEDICAL SYSTEMS CORP. | PCF-Q180AI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |