EVIS EXERA II COLONOVIDEOSCOPE

MDR report key: 6857326 · Received September 12, 2017

Report

Report Number
8010047-2017-01329
Event Type
Malfunction
Date Received
September 12, 2017
Date of Event
August 11, 2017
Report Date
September 12, 2017
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDF
PMA / PMN Number
PK100584
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. BUT WAS RETURNED TO OLYMPUS (B)(4). FOLLOWING ADDITIONAL HIGH LEVEL DISINFECTION AT (B)(4), THE SUBJECT DEVICE WAS SENT TO A THIRD PARTY LABORATORY FOR ADDITIONAL MICROBIOLOGICAL TESTING. IN THE ADDITIONAL TEST, THE TESTING INDICATED NO MICROBIAL GROWTH FOR THE SUBJECT DEVICE. OMSC REVIEWED THE MANUFACTURING HISTORY OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING ROUTINE SURVEILLANCE CULTURING TEST BY THE FACILITY, THE SUBJECT DEVICE TESTED POSITIVE FOR BACTERIA. THE CHANNELS OF THE SUBJECT DEVICE EXCEPT FOR THE AUXILIARY CHANNEL TESTED POSITIVE FOR FOLLOWING CFU OF BACTERIA. (B)(6). THE CUSTOMER REPORTED THAT THE SUBJECT DEVICE WAS REPROCESSED ACCORDING TO THE INSTRUCTION FOR USE AND THE FRENCH REGULATION. THE SUBJECT DEVICE HAD BEEN DISINFECTED USING AN OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR MODEL ETD-4 (NOT AVAILABLE IN THE U.S.) WITH PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.