FDA Adverse Event Malfunction Summary report: N

I-STAT EG7+ CARTRIDGE

MDR report key: 685621 · Received March 6, 2006

Report

Report Number
2245578-2006-00003
Event Type
Malfunction
Date Received
March 6, 2006
Date of Event
February 6, 2006
Report Date
March 6, 2006
Manufacturer
I-STAT CORP.
Product Code
MMI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

CUSTOMER REPORTED IN 2/2006 THAT A PATIENT SAMPLE GENERATED AN I-STAT HEMATOCRIT OF 25% PCV AND THE LABORATORY RESULT WAS 29% PCV. A SPUN HEMATOCRIT WAS PERFORMED ON THE SAMPLE WHICH GENERATED A RESULT OF 29% PCV. NO TREATMENT WAS GIVEN BASED ON THE HEMATOCRIT RESULT OF 25% PCV.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT EG7+ CARTRIDGE IVD MMI I-STAT CORP. NA U05249A

Patients

Seq Age Sex Outcome Treatment
1 79 YR