FDA Adverse Event Malfunction Summary report: N

SURGICAL CONVENIENCE KIT

MDR report key: 6854630 · Received September 11, 2017

Report

Report Number
0001055892-2017-00010
Event Type
Malfunction
Date Received
September 11, 2017
Date of Event
July 25, 2017
Report Date
November 3, 2017
Manufacturer
SYNERGY HEALTH NORTH AMERICA
Product Code
LRO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

NO DELAYS OR CANCELLATIONS IN PATIENT PROCEDURE WERE REPORTED. THE USER WAS NOT WEARING THE APPROPRIATE GOWN FOR THE PROCEDURE WHEN THE REPORTED EVENT OCCURRED. THE USER WAS WEARING A LEVEL 2 GOWN AND SHOULD HAVE BEEN WEARING A LEVEL 3 OR LEVEL 4 GOWN TO PROVIDE ADEQUATE BARRIER PROTECTION. SYNERGY HEALTH NORTH AMERICA ACCOUNT MANAGER WILL BE CONDUCTING IN-SERVICE TRAINING WITH THE USER FACILITY ON THE PROPER USAGE OF THE GOWNS.

Additional Manufacturer Narrative · 1

SYNERGY HEALTH NORTH AMERICA REPRESENTATIVE CONDUCTED CUSTOMER (WAKEMED) IN-SERVICE TRAINING ON OCTOBER 06, 2017 AND OCTOBER 09, 2017 ON THE PROPER USAGE OF SURGICAL GOWNS.

Additional Manufacturer Narrative · 1

SYNERGY HEALTH NORTH AMERICA REPRESENTATIVE CONDUCTED CUSTOMER (WAKEMED) IN-SERVICE TRAINING ON OCTOBER 06, 2017 AND OCTOBER 09, 2017 ON THE PROPER USAGE OF SURGICAL GOWNS.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT AT THE END OF A SURGICAL PROCEDURE AN EMPLOYEE OBSERVED A STRIKE-THROUGH ON THEIR SURGICAL GOWN.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT THE END OF A SURGICAL PROCEDURE AN EMPLOYEE OBSERVED A STRIKE-THROUGH ON THEIR SURGICAL GOWN.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT AT THE END OF A SURGICAL PROCEDURE AN EMPLOYEE OBSERVED A STRIKETHROUGH ON THEIR SURGICAL GOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
634617 SURGICAL CONVENIENCE KIT SURGICAL GOWN LRO SYNERGY HEALTH NORTH AMERICA 1003D

Patients

Seq Age Sex Outcome Treatment
1 Other