FDA Adverse Event Malfunction Summary report: N

60 ML BD¿ SYRINGE WITH BD LUER-LOK¿ TIP

MDR report key: 6853356 · Received September 8, 2017

Report

Report Number
1911916-2017-00215
Event Type
Malfunction
Date Received
September 8, 2017
Date of Event
August 16, 2017
Report Date
November 20, 2017
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
PMA / PMN Number
K110771
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: SAMPLE WAS RECEIVED IN THE PLANT FOR EVALUATION. BREAKOUT SUSTAIN WAS PERFORMED ON PART FROM BD MEDICAL AND WAS IN SPEC. THIS IS AN UNCONFIRMED DEFECT. DHR: PRODUCT WAS RAN ON 860 #2 LINE ON 5/6/2017. AT BLISTERPAK THERE WERE 18 INSPECTIONS ON 144 PARTS WITH ZERO DEFECTS FOUND. AT PRINT/ASSY THERE WERE 18 INSPECTIONS ON 900 PARTS WITH ZERO DEFECTS FOUND. INVESTIGATION CONCLUSION: BREAKOUT SUSTAIN WAS PERFORMED ON BD MEDICAL SYRINGE SPEC: BREAKOUT-0.0-9.0LBS SUSTAIN-0.0-5.0LBS. READING WAS BREAKOUT-1.78LBS AND SUSTAIN-.96LBS. COMPLAINT TRENDING REVIEW OF THIS LOT AND OR PRODUCT FAMILY FOR THIS ISSUE REVEALS ONE COMPLAINT. SILICONE WEIGHT TESTING IS PERFORMED TWICE DURING THE BATCH (MID AND END). SPEC: .0055-.0120GRAMS. READINGS WERE MIN. FOR SILICONE.0075G AND MAX WAS .0096G. ROOT CAUSE DESCRIPTION: BREAKOUT SUSTAIN WAS PERFORMED ON PART FROM BD MEDICAL AND WAS IN SPEC. THIS IS A UNCONFIRMED DEFECT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PLUNGER MOVEMENT WAS DIFFICULT ON A 60 ML BD¿ SYRINGE WITH BD LUER-LOK¿ TIP DURING USE. NO INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
633259 60 ML BD¿ SYRINGE WITH BD LUER-LOK¿ TIP SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 7086739

Patients

Seq Age Sex Outcome Treatment
1 Other