SSCOR
Report
- Report Number
- 2022724-2017-00002
- Event Type
- Death
- Date Received
- September 8, 2017
- Date of Event
- April 19, 2017
- Report Date
- May 17, 2017
- Manufacturer
- SSCOR, INC
- Product Code
- BTA
- PMA / PMN Number
- K945929
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ON (B)(6) 2017 SSCOR RECEIVED AN EMAIL FROM (B)(6), (B)(6) OF (B)(6) EMS. THE EMAIL INDICATED IN THE PAST WEEK THE DEPARTMENT HAD TWO SSCOR VX-2 (MODEL 2310BV) SUCTION UNITS STOP WORKING DURING PATIENT CARE SITUATIONS. AFTER THE EMAIL WAS RECEIVED SSCOR FOLLOWED UP WITH (B)(6) TO GATHER ADDITIONAL INFORMATION. THE INITIAL INDICATIONS WERE THE DEVICE STOPPED WORKING BECAUSE THE INTERNAL BATTERY IN THE SUCTION DEVICE HAD LOST CAPACITY. SSCOR REQUESTED THE DEPARTMENT CHECK ALL OF THE BATTERIES IN THEIR DEVICES BY CONDUCTING THE BATTERY TEST AS DESCRIBED IN THE INSTRUCTIONS FOR USE FOR THE DEVICES. ADDITIONALLY, SSCOR SENT OUT A TOTAL OF (B)(4) NEW REPLACEMENT BATTERIES FOR ALL THE SSCOR SUCTION UNITS OWNED BY THE DEPARTMENT AS A PRECAUTIONARY MEASURE. ON (B)(6) 2017 THE DEPARTMENT REPORTED THEY FOUND THAT (B)(4) OF THE DEVICES HAD BATTERIES WITH LOW CAPACITY. THEY WERE ABLE TO REPLACE THESE LOW CAPACITY BATTERIES WITH THE REPLACEMENT BATTERIES SSCOR SENT OUT. SSCOR ALSO DISCUSSED THE BATTERY CARE AND MAINTENANCE PROCEDURES AS DESCRIBED IN THE OPERATIONS MANUAL WITH THE DEPARTMENT. THE DEPARTMENT WAS NOT FOLLOWING THE RECOMMENDED BATTERY MAINTENANCE PROCEDURES. AFTER THE DISCUSSION THE DEPARTMENT AGREED TO UPDATE THEIR INTERNAL PROTOCOLS AND FOLLOW THE BATTERY CARE AND MAINTENANCE PROCEDURES AS DESCRIBED IN THE OPERATIONS MANUAL FOR THE DEVICES.
PARAMEDICS WERE CALLED TO AID A PATIENT WITH BREATHING PROBLEMS. THE PARAMEDICS ARRIVED TO FIND THE PATIENT UNRESPONSIVE, PULSE-LESS AND APNEIC. ADDITIONALLY THE PATIENTS PUPILS WERE FIXED AND NON-REACTIVE AND MOTOR SKILLS WERE FOUND TO BE FLACCID. THE PARAMEDICS INITIATED CPR ACTIVITIES AND ATTEMPTED TO SUCTION THE PATIENT AT THE SCENE USING A SSCOR SUCTION DEVICE. THE DEVICE WOULD NOT OPERATE AT THAT TIME. THE PATIENT WAS TRANSPORTED TO THE VEHICLE AND THE SUCTION UNIT WAS INSTALLED INTO ITS CHARGING / RETENTION BRACKET AT WHICH TIME THE DEVICE OPERATED TO SPECIFICATIONS AS IT WAS BEING POWERED BY THE VEHICLE BATTERY. THE PATIENT WAS TRANSPORTED TO THE HOSPITAL WHERE HE WAS PRONOUNCED DEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631226 | SSCOR | VX-2 | BTA | SSCOR, INC | 2310BV | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Death |