FDA Adverse Event Injury Summary report: N

AMISTEM COLLARED STEM, HA COATED SIZE 6 STD

MDR report key: 6851066 · Received September 8, 2017

Report

Report Number
3005180920-2017-00505
Event Type
Injury
Date Received
September 8, 2017
Date of Event
August 10, 2017
Report Date
September 8, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804618
PMA / PMN Number
K121011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 25 AUGUST 2017 THE MEDICAL AFFAIRS PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: LATE INFECTION IN CEMENTLESS THA, 2 YEARS AFTER PRIMARY OPERATION. INFECTION IS A KNOWN POSSIBLE ADVERSE EVENT FOLLOWING EVERY SURGERY, INCLUDING THA'S. TO DATE, THERE IS NO REASON TO SUSPECT THAT THE CAUSE MAY BE LINKED TO THE IMPLANTED DEVICES. BATCH REVIEWS PERFORMED ON 04 SEPTEMBER 2017. LOT 146198: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13 JANUARY 2015. EXPIRATION DATE: 2019-11-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE S -3.5, CODE 01.29.201, LOT. 151720 (K112115) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 29 SEPTEMBER 2015. EXPIRATION DATE: 2020-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. VERSAFITCUP ACETABULAR SHELL Ø 54, CODE 01.26.54MB, LOT. 152020 (K083116) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26 AUGUST 2015. EXPIRATION DATE: 2020-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. VERSAFITCUP DOUBLE MOBILITY HC LINER Ø 54/28, CODE 01.26.2854MHC, LOT. 152103 (K092265) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24 AUGUST 2015. EXPIRATION DATE: 2020-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN DUE TO SIGNS OF INFECTION (LEFT HIP). THE PATHOGEN IS UNKNOWN. THE INFECTION CAUSED THE STEM TO BECOME LOOSE. THE SURGEON REVISED ALL MEDACTA HARDWARE AND RE-IMPLANTED PERMANENT PRODUCT. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS ARE AVAILABLE. EXPLANTS ARE NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631449 AMISTEM COLLARED STEM, HA COATED SIZE 6 STD CEMENTLESS FEMORAL STEM LZO MEDACTA INTERNATIONAL SA 146198 07630030804618

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention