GMK-HINGE FEMORAL COMPONENT SIZE 2 RIGHT
Report
- Report Number
- 3005180920-2017-00502
- Event Type
- Injury
- Date Received
- September 8, 2017
- Date of Event
- August 11, 2017
- Report Date
- September 8, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- KRO
- UDI-DI
- 07630030825422
- PMA / PMN Number
- K130299
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
BATCH REVIEWS PERFORMED ON (B)(6) 2017. LOT 162147: 17 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2016. EXPIRATION DATE: (B)(6) 2021. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 7 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK-HINGE EXTENSION STEM - FLUTED Ø 16 L 65, CODE 02.07.FCL16065, LOT. 162547 (K120790) 40 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2016. EXPIRATION DATE: (B)(6) 2021. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 21 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK-HINGE FIXED TIBIAL INSERT SIZE 2/12MM, CODE 02.09.0212H, LOT. 150741 (K130299) 20 INSERTS MANUFACTURED AND RELEASED ON (B)(6) 2015. EXPIRATION DATE: (B)(6) 2020. NO ANOMALIES FOUND RELATED TO THE ISSUE. TO DATE, 13 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
THE PATIENT CAME IN COMPLAINING OF LIMITED RANGE OF MOTION. THE PATIENT HAD 40 DEGREES CONTRACTURE FOLLOWING THE PRIMARY TKA WITH A GMK-HINGE. PREOPERATIVE, THE PATIENT WAS 25 DEGREES CONTRACTURE. THE SURGEON PERFORMED A FEMORAL SHORTENING, WHICH ALLOWED PATIENT TO HAVE FULL RANGE OF MOTION, AND REVISED THE FEMUR, POLY AND STEM. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AND EXPLANTS ARE NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632418 | GMK-HINGE FEMORAL COMPONENT SIZE 2 RIGHT | KNEE FEMORAL COMPONENT | KRO | MEDACTA INTERNATIONAL SA | 162147 | 07630030825422 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |