FDA Adverse Event
Other
Summary report: N
BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 684909
·
Received March 7, 2006
Report
- Report Number
- 2243072-2006-00101
- Event Type
- Other
- Date Received
- March 7, 2006
- Date of Event
- February 9, 2006
- Report Date
- March 6, 2006
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED WITH ACCURACY ISSUE. BELIEVES HER CHILD IS GETTING HIGHER READINGS THAN THE WAY SHE FEELS. ANALYSIS RUN ON CONSENSUS ERROR GRID (CEG) USING COMPETITIVE MEAN READINGS (49) AS REFERENCE POINT VS. BD READINGS (200, 269) YIELDED SOME DATA POINTS IN THE D RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |