FDA Adverse Event Malfunction Summary report: N

3.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM

MDR report key: 6836452 · Received August 31, 2017

Report

Report Number
9612488-2017-10451
Event Type
Malfunction
Date Received
August 31, 2017
Date of Event
August 10, 2017
Report Date
August 10, 2017
Manufacturer
SYNTHES BETTLACH
Product Code
HTW
UDI-DI
07611819775398
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A PRODUCT INVESTIGATION WAS PERFORMED. THE DRILL BITS (PART # 03.010.103, LOT # 8735037 & LOT # 8393880) WERE RECEIVED FOR INVESTIGATION. THE INVESTIGATION HAS SHOWN THAT BOTH DRILL BIT TIPS WERE FOUND BROKEN. FURTHERMORE, THE CUTTING EDGES ARE BLUNT AND THERE ARE NICKS AND WEAR AND TEAR SIGNS ON SURFACE. THE BROKEN PARTS ARE MISSING. THE MANUFACTURING REVIEW SHOWS THAT THE PRODUCTION PROCEDURE WAS ACCORDING TO THE SPECIFICATIONS AND THERE WERE NO ISSUES THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. AS INDICATED IN THE MANUFACTURING DOCUMENTS FOR LOT 8735037, THE CORRECT MATERIAL WAS USED AND THE HARDNESS WAS WITH THE MEASUREMENT OF 54.5 HRC AND 54.8 HRC WITHIN THE TOLERANCE OF A MINIMUM OF 52 HRC AND MAXIMUM 55 HRC. AS INDICATED IN THE MANUFACTURING DOCUMENTS FOR LOT 8393880, THE CORRECT MATERIAL WAS USED AND THE HARDNESS WAS WITH THE MEASUREMENT OF 52.5 HRC AND 53.0 HRC WITHIN THE TOLERANCE OF A MINIMUM OF 52 HRC AND MAXIMUM 55 HRC. THE MICROSCOPIC VIEW, WITH A MAGNIFICATION OF 10X, OF THE BROKEN SURFACE SHOWS A HOMOGENOUS SURFACE WHAT INDICATES MATERIAL CONFORMITY AS WELL. BECAUSE OF THE DAMAGE, THE COMPLAINT RELEVANT DIMENSIONS CANNOT BE CHECKED FOR DIMENSIONAL ACCURACY OF THE VALID MANUFACTURING SPECIFICATIONS. WE ARE NOT ABLE TO DETERMINE THE EXACT CAUSE OF THIS OCCURRENCE DUE TO MISSING DETAILED CLINICAL INFORMATION AND MISSING BROKEN PART. WE DO SUPPOSE THAT A MECHANICAL OVERLOAD SITUATION DURING USE COULD LEAD TO THE BREAKAGE. UNFORTUNATELY, WE ONLY HAVE LIMITED INFORMATION IN THE COMPLAINT DESCRIPTION AND CANNOT CONFIRM HOW THIS HAPPENED. BASED ON THE INVESTIGATION RESULTS WE CONCLUDE THAT THE CAUSE OF FAILURE IS NOT DUE TO ANY MANUFACTURING NON-CONFORMANCES NO PRODUCT FAULT COULD BE DETECTED. CORRECTED DATA: DATE OF EVENT WAS REPORTED AS (B)(6) 2017; HOWEVER, IT IS UNKNOWN IF THAT IS THE DATE THE DEVICE BROKE OR THE DATE IT WAS DISCOVERED BROKEN. REPORTER ADDRESS. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT, NOT DIAGNOSIS. NO PATIENT INVOLVEMENT REPORTED. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. DATE RETURNED TO MANUFACTURER. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ONGOING; NO CONCLUSION COULD BE DRAWN AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. REPORTERS PHONE NUMBER: (B)(6). DEVICE HISTORY RECORDS REVIEW WAS CONDUCTED. THE REPORT INDICATES THAT THE: PART #03.010.103/ LOT#8393880. MANUFACTURING LOCATION: (B)(4). MANUFACTURING DATE: 30.APR.2013. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES ON AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE OBTAINED LOAN SET FROM THE DISTRIBUTOR (B)(4) AFTER CHECKING CONTAINS DEFECTIVE DEVICES: DRILL BIT 03.010.103 (BOTH) IS THELATHE TIP IS BROKEN. DRILL BIT 03.019.016 (BOTH) IS DAMAGE TO THE DRILL, IT IS ASSUMED THAT THESE DRILLS THEMSELVES HAVE BEEN SHARPENED. NO SURGICAL DELAY IS REPORTED. THERE IS NO PATIENT INVOLVED. COMPLAINT INVOLVES 4 PART. THIS REPORT IS 2 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
615424 3.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM BIT,DRILL HTW SYNTHES BETTLACH 8393880 07611819775398

Patients

Seq Age Sex Outcome Treatment
1