FDA Adverse Event Malfunction Summary report: N

HIP END EFFECTOR, VARIABLE ANGLE

MDR report key: 6835009 · Received August 31, 2017

Report

Report Number
3005985723-2017-00413
Event Type
Malfunction
Date Received
August 31, 2017
Date of Event
August 10, 2017
Report Date
October 23, 2017
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K170593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

REPORTED EVENT: HIP END EFFECTOR BROKE DURING IMPACTION. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION: VISUAL INSPECTION CONFIRMS FAILED LOCKING MECHANISM WITH BALL BEARINGS FAILING. DISCOLORATION AND WEAR ARE SEEN ON THE 202862 HIP BALL RETAINER. VISUAL INSPECTION CONFIRMS A SHEARED OFF 202866 HIP RELEASE KNOB. FRACTURE IS SEEN AT THE DEGREE MARKERS OF THE 20966 REAMER BARREL. SEE ATTACHMENTS FOR IMAGES OF THE INSTRUMENT. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED AS VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION AS NOT COMPLETED AS VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED UNDER LOT NO 19030415 AND (B)(4) INCLUDING S/N (B)(4) ACCEPTED INTO FINAL STOCK ON 12/13/2014 . A REVIEW OF (B)(4). COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 206967, LOT 19030415 SHOWS 6 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. (B)(4). CONCLUSIONS: ALLEGED FAILURE CONFIRMED. CORRECTIVE ACTION/PREVENTIVE ACTION: CAPA (B)(4) WAS INITIATED BASED ON OTHER COMPLAINTS REPORTING THIS ISSUE.

Description of Event or Problem · 1

HIP END EFFECTOR BROKE DURING IMPACTION.

Description of Event or Problem · 1

HIP END EFFECTOR BROKE DURING IMPACTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
615586 HIP END EFFECTOR, VARIABLE ANGLE STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 19030415

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization