LAP CHOLE PACK
Report
- Report Number
- 3005011024-2017-00013
- Event Type
- Malfunction
- Date Received
- August 31, 2017
- Date of Event
- August 10, 2017
- Report Date
- February 21, 2018
- Manufacturer
- DEROYAL INDUSTRIES, INC.
- Product Code
- FDE
- PMA / PMN Number
- ENFORCEMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
ROOT CAUSE: THE DRAPE CONTAINED WITHIN THE FINISHED GOOD KIT (89-6201) IS SUPPLIED TO DEROYAL BY WELMED. A SUPPLIER CORRECTIVE ACTION REQUEST (SCAR) WAS ISSUED TO WELMED. THE VENDOR RESPONDED INDICATING THE FOAM REINFORCEMENT MATERIAL AROUND THE FENESTRATION NEEDED TO BE CHANGED. IT HAS CHANGED THIS MATERIAL TO A MORE DURABLE FIBER. CORRECTIVE ACTION: IN ITS SCAR RESPONSE, WELMED STATED IT IS CHANGING ITS SPECIFICATIONS FOR THE FOAM REINFORCEMENT MATERIAL. WELMED HAS DOCUMENTED THIS ISSUE FOR TRENDING AND WILL CONTINUE TO MONITOR THE SITUATION IN ITS CAPA PROGRAM. INVESTIGATION SUMMARY: AN INTERNAL COMPLAINT (CALL 41751) WAS RECEIVED INDICATING THAT A LAP CHOLE PACK (FINISHED GOOD 89-6201, LOT 42061936) CONTAINED A FENESTRATED KEY DRAPE THAT WAS FLAKING, ALLOWING DEBRIS INTO THE SURGICAL SITE. THE SAMPLE WAS RETURNED FOR EVALUATION. DEROYAL QUALITY CONTROL PERSONNEL EVALUATED THE SAMPLE AND CONFIRMED THE DRAPE BORDER WAS FLAKING. THE WORK ORDER FOR THE REPORTED LOT NUMBER WAS REVIEWED FOR DISCREPANCIES THAT MAY HAVE CONTRIBUTED TO THE REPORTED EVENT. NO DISCREPANCIES WERE IDENTIFIED. AN INVENTORY CHECK COULD NOT BE COMPLETED BECAUSE DEROYAL NO LONGER KEEPS THIS COMPONENT (5-18466-D) ON HAND. THE LAST FINISHED GOOD PRODUCED WITH THIS RAW MATERIAL WAS MANUFACTURED IN JUNE 2016. THE AFFECTED DRAPE (RAW MATERIAL 5-18466-D) WAS SUPPLIED TO DEROYAL BY WELMED. A SCAR WAS SENT TO WELMED AND A RESPONSE HAS BEEN RECEIVED. THE 2014-2016 SCAR AND SUPPLIER NOTIFICATION LETTER LOGS WERE REVIEWED FOR SIMILAR COMPLAINTS. SIMILAR COMPLAINTS WERE IDENTIFIED. PREVENTIVE ACTION: A PREVENTIVE ACTION HAS NOT BEEN TAKEN. THE INVESTIGATION IS COMPLETE AT THIS TIME. IF NEW AND CRITICAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.
INVESTIGATION SUMMARY AN INTERNAL COMPLAINT (CALL (B)(4)) WAS RECEIVED INDICATING THAT A LAP CHOLE PACK (FINISHED GOOD 89-6201, LOT 42061936) CONTAINED A FENESTRATED KEY DRAPE THAT WAS FLAKING, ALLOWING DEBRIS INTO THE SURGICAL SITE. THE SAMPLE IS REPORTED TO BE AVAILABLE FOR RETURN. HOWEVER, AS OF THE DATE OF THIS REPORT, THE SAMPLE HAD NOT BEEN RECEIVED FOR EVALUATION. THE WORK ORDER FOR THE REPORTED LOT NUMBER WAS REVIEWED FOR DISCREPANCIES THAT MAY HAVE CONTRIBUTED TO THE REPORTED EVENT. NO DISCREPANCIES WERE IDENTIFIED. THE AFFECTED DRAPE (RAW MATERIAL (B)(4)) WAS SUPPLIED TO DEROYAL BY (B)(4). A SUPPLIER CORRECTIVE ACTION REQUEST (B)(4) WAS ISSUED AUGUST 15, 2017, AND A RESPONSE IS DUE SEPTEMBER 26, 2017. THE INVESTIGATION IS ONGOING AT THIS TIME. WHEN NEW AND CRITICAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.
THE FENESTRATED KEY SHEET WAS FLAKING, WHICH LED TO DEBRIS IN THE SURGICAL SITE.
THE FENESTRATED KEY SHEET WAS FLAKING, WHICH LED TO DEBRIS IN THE SURGICAL SITE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 614820 | LAP CHOLE PACK | LAPAROSCOPY KIT | FDE | DEROYAL INDUSTRIES, INC. | 89-6201 | 42061936 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |