FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 6823765 · Received August 28, 2017

Report

Report Number
2531779-2017-19416
Event Type
Malfunction
Date Received
August 28, 2017
Report Date
August 2, 2017
Manufacturer
ANIMAS CORPORATION
Product Code
OYC
UDI-DI
10840406100037
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/02/2017 WITH THE FOLLOWING FINDINGS: THE BATTERY COMPARTMENT WAS CRACKED BELOW THE BUMPER PAD. INITIAL REPORTER: ANIMAS CORPORATION. (B)(4).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE BATTERY COMPARTMENT WAS CRACKED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 08/02/2017.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
606027 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS CORPORATION 10840406100037

Patients

Seq Age Sex Outcome Treatment
1 40 YR