FDA Adverse Event Malfunction Summary report: N

I-STAT1 ANALYZER, IMMUNO READY

MDR report key: 6819961 · Received August 25, 2017

Report

Report Number
2245578-2017-00278
Event Type
Malfunction
Date Received
August 25, 2017
Date of Event
August 9, 2017
Report Date
November 17, 2017
Manufacturer
ABBOTT POINT OF CARE
Product Code
CGA
UDI-DI
00054749001408
PMA / PMN Number
K103195
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 11/07/2017. THE FAILURE WAS DUE TO A DEFECTIVE TANTALUM CAPACITOR.

Description of Event or Problem · 1

ON (B)(6)2017, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT ANALYZER SN (B)(4) WAS FEELING "EXTREMELY HOT" AND NOT ACTIVATING . THE CUSTOMER ALSO STATED THAT THERE WAS NO INDICATION OF SMOKE, OR BURNING SMELL. THERE WAS NO IMPACT TO PATIENT CARE OR MANAGEMENT AS PATIENTS WERE TESTED ON OTHER I-STAT ANALYZERS. THE I-STAT WAS OPERATING WITH RECHARGEABLE BATTERIES. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

Description of Event or Problem · 1

NA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602019 I-STAT1 ANALYZER, IMMUNO READY I-STAT1 ANALYZER CGA ABBOTT POINT OF CARE NA NA 00054749001408

Patients

Seq Age Sex Outcome Treatment
1