FDA Adverse Event Malfunction Summary report: N

20 G X 1.00 IN. (1.1 MM X 25 MM) BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER

MDR report key: 6815706 · Received August 23, 2017

Report

Report Number
1710034-2017-00152
Event Type
Malfunction
Date Received
August 23, 2017
Date of Event
July 1, 2017
Report Date
October 6, 2017
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
PMA / PMN Number
K110443
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Additional Manufacturer Narrative · 1

INVESTIGATION: DHR REVIEW WAS PERFORMED ON THE FOLLOWING LOT NUMBER: 6181671. THE LOT NUMBER WAS BUILT ON AFA MANUFACTURING/PACKAGING LINE 11 FROM JULY 1, 2016 THRU JULY 6, 2017. PER REVIEW OF THE DHR IT WAS CONCLUDED THAT ALL REQUIRED CHALLENGES SAMPLES AND TESTING WAS PERFORMED PER SPECIFICATION S-AU17 IN ACCORDANCE WITH THE SET ¿UP AND IN-PROCESS SAMPLING PLANS. PER REVIEW IT WAS NOTED THAT THERE WERE NO REJECT ACTIVITY FINDINGS THROUGHOUT THE BUILD OF THIS LOT THAT WOULD IMPACT UPON THE QUALITY OF THE PRODUCT. OBSERVATIONS AND TESTING COULD NOT BE PERFORMED BECAUSE UNITS WERE NOT RECEIVED FOR INVESTIGATION OF THIS INCIDENT. CONCLUSIONS: THE DEFECT OF PACKAGE SEAL INTEGRITY POOR/QUESTIONABLE, STATED IN THE DESCRIPTION OF THE COMPLAINT, COULD NOT BE IDENTIFIED OR CONFIRMED AND CAUSE COULD NOT BE DETERMINED AS THE UNITS DESCRIBED IN THE PRODUCT INCIDENT REPORT WERE NOT RETURNED FOR EVALUATION AND TESTING. THEREFORE, THERE WAS NO PHYSICAL EVIDENCE TO CONFIRM OR TO SUPPORT MANUFACTURING PROCESS RELATED ISSUES FOR THE REPORTED DEFECT. SAMPLES WERE NOT RETURNED FOR EVALUATION AND THERE WAS NO PHYSICAL EVIDENCE TO CONFIRM OR TO SUPPORT MANUFACTURING PROCESS RELATED ISSUES FOR THE DEFECT. DUE TO EVALUATION OF OTHER COMPLAINT SAMPLES, CAPA (B)(4) HAS BEEN OPENED TO INVESTIGATE THE PACKAGE OPEN SEAL DEFECTS AND IMPLEMENT CORRECTIVE ACTIONS.

Description of Event or Problem · 1

IT WAS REPORTED THE PACKAGING ON A 20 G X 1.00 IN. (1.1 MM X 25 MM) BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WAS OPEN BEFORE USE. NO MEDICAL INTERVENTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
594663 20 G X 1.00 IN. (1.1 MM X 25 MM) BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 6181671

Patients

Seq Age Sex Outcome Treatment
1 Other