20 G X 1.00 IN. (1.1 MM X 25 MM) BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER
Report
- Report Number
- 1710034-2017-00152
- Event Type
- Malfunction
- Date Received
- August 23, 2017
- Date of Event
- July 1, 2017
- Report Date
- October 6, 2017
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- PMA / PMN Number
- K110443
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
INVESTIGATION: DHR REVIEW WAS PERFORMED ON THE FOLLOWING LOT NUMBER: 6181671. THE LOT NUMBER WAS BUILT ON AFA MANUFACTURING/PACKAGING LINE 11 FROM JULY 1, 2016 THRU JULY 6, 2017. PER REVIEW OF THE DHR IT WAS CONCLUDED THAT ALL REQUIRED CHALLENGES SAMPLES AND TESTING WAS PERFORMED PER SPECIFICATION S-AU17 IN ACCORDANCE WITH THE SET ¿UP AND IN-PROCESS SAMPLING PLANS. PER REVIEW IT WAS NOTED THAT THERE WERE NO REJECT ACTIVITY FINDINGS THROUGHOUT THE BUILD OF THIS LOT THAT WOULD IMPACT UPON THE QUALITY OF THE PRODUCT. OBSERVATIONS AND TESTING COULD NOT BE PERFORMED BECAUSE UNITS WERE NOT RECEIVED FOR INVESTIGATION OF THIS INCIDENT. CONCLUSIONS: THE DEFECT OF PACKAGE SEAL INTEGRITY POOR/QUESTIONABLE, STATED IN THE DESCRIPTION OF THE COMPLAINT, COULD NOT BE IDENTIFIED OR CONFIRMED AND CAUSE COULD NOT BE DETERMINED AS THE UNITS DESCRIBED IN THE PRODUCT INCIDENT REPORT WERE NOT RETURNED FOR EVALUATION AND TESTING. THEREFORE, THERE WAS NO PHYSICAL EVIDENCE TO CONFIRM OR TO SUPPORT MANUFACTURING PROCESS RELATED ISSUES FOR THE REPORTED DEFECT. SAMPLES WERE NOT RETURNED FOR EVALUATION AND THERE WAS NO PHYSICAL EVIDENCE TO CONFIRM OR TO SUPPORT MANUFACTURING PROCESS RELATED ISSUES FOR THE DEFECT. DUE TO EVALUATION OF OTHER COMPLAINT SAMPLES, CAPA (B)(4) HAS BEEN OPENED TO INVESTIGATE THE PACKAGE OPEN SEAL DEFECTS AND IMPLEMENT CORRECTIVE ACTIONS.
IT WAS REPORTED THE PACKAGING ON A 20 G X 1.00 IN. (1.1 MM X 25 MM) BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER WAS OPEN BEFORE USE. NO MEDICAL INTERVENTIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 594663 | 20 G X 1.00 IN. (1.1 MM X 25 MM) BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 6181671 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |