NITINOL TC ELECTRODE
Report
- Report Number
- 3006630150-2017-03071
- Event Type
- Malfunction
- Date Received
- August 23, 2017
- Date of Event
- August 1, 2017
- Report Date
- August 1, 2017
- Manufacturer
- COSMAN MEDICAL
- Product Code
- GXI
- UDI-DI
- 00813250010107
- PMA / PMN Number
- K050084
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
EXPIRATION DATE: NA MODEL #: CSK-TC10 LOT # K237 DESCRIPTION: CSK TC ELECTRODE, 10CM QUANTITY: 1 MODEL #: CSK-TC10 LOT # K509 DESCRIPTION: CSK TC ELECTRODE, 10CM QUANTITY: 7 MODEL #: CSK-TC10 LOT # K239 DESCRIPTION: CSK TC ELECTRODE, 10CM QUANTITY: 4 MODEL #: CSK-TC10 LOT # L225 DESCRIPTION: CSK TC ELECTRODE, 10CM QUANTITY: 1 MODEL #: CSK-TC15 LOT # K240 DESCRIPTION: CSK TC ELECTRODE, 15 CM QUANTITY: 1 MODEL #: CSK-TC15 LOT #: 022317 DESCRIPTION: CSK TC ELECTRODE, 15 CM QUANTITY: 1 MODEL #: CSK-TC5 LOT # 031017 DESCRIPTION: CSK TC ELECTRODE, 5 CM QUANTITY: 1.
MODEL #: CSK-TC10 LOT # K237 ¿ QUANTITY 1 LOT # K509 ¿ QUANTITY 10 LOT # K239 ¿ QUANTITY 7 LOT # L225 ¿ QUANTITY 1 DESCRIPTION: CSK TC ELECTRODE, 10CM MODEL #: CSK-TC15 LOT # K240 ¿ QUANTITY 2 LOT #: 022317 ¿ QUANTITY 1 DESCRIPTION: CSK TC ELECTRODE, 15 CM MODEL #: CSK-TC5 LOT # 03102017 ¿ QUANTITY 1 DESCRIPTION: CSK TC ELECTRODE, 5 CM THE TWENTY-THREE RETURNED ELECTRODES WERE ANALYZED AND EXTERNAL VISUAL INSPECTION REVEALED TRACES OF POTTING EPOXY ON ALL THE HUBS. THE EPOXY LEAKED THROUGH THE SHAFT OPENING IN ITS LIQUID STATE DURING THE MANUFACTURING PROCESS, AND WAS NOT REMOVED AFTER CURING.
A REPORT WAS RECEIVED THAT AFTER THE ELECTRODES WERE CLEANED AND STERILIZED, IT WAS NOTED THAT THE CABLES HAD MELTED EPOXY THAT WAS BUBBLING OUT FROM THE BASE OF THE NEEDLES.
A REPORT WAS RECEIVED THAT AFTER THE ELECTRODES WERE CLEANED AND STERILIZED, IT WAS NOTED THAT THE CABLES HAD MELTED EPOXY THAT WAS BUBBLING OUT FROM THE BASE OF THE NEEDLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 592941 | NITINOL TC ELECTRODE | ELECTRODE | GXI | COSMAN MEDICAL | CSK-TC10 | L225 | 00813250010107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |