FDA Adverse Event
Malfunction
Summary report: N
CARESTATION 650
MDR report key: 6810539
·
Received August 22, 2017
Report
- Report Number
- 9710602-2017-00204
- Event Type
- Malfunction
- Date Received
- August 22, 2017
- Date of Event
- August 1, 2017
- Report Date
- August 22, 2017
- Manufacturer
- GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
- Product Code
- BSZ
- PMA / PMN Number
- K151570
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). DURING GE HEALTHCARE SYSTEM CHECKOUT, IT WAS NOTED THAT THE DISPOSABLE CO2 CANISTER (LOT 9170217) WAS APPROXIMATELY 1 MM SHORTER IN HEIGHT THAN A REUSABLE CANISTER. REPLACED CUSTOMER SUPPLIED DISPOSABLE CO2 CANISTER FROM LOT 3160517 TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT, DURING PATIENT USE, THE CO2 LEVEL WENT UP. THERE WAS NO REPORTED PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 591631 | CARESTATION 650 | ANESTHESIA GAS MACHINE | BSZ | GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |