FDA Adverse Event Malfunction Summary report: N

CARESTATION 650

MDR report key: 6810539 · Received August 22, 2017

Report

Report Number
9710602-2017-00204
Event Type
Malfunction
Date Received
August 22, 2017
Date of Event
August 1, 2017
Report Date
August 22, 2017
Manufacturer
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
Product Code
BSZ
PMA / PMN Number
K151570
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING GE HEALTHCARE SYSTEM CHECKOUT, IT WAS NOTED THAT THE DISPOSABLE CO2 CANISTER (LOT 9170217) WAS APPROXIMATELY 1 MM SHORTER IN HEIGHT THAN A REUSABLE CANISTER. REPLACED CUSTOMER SUPPLIED DISPOSABLE CO2 CANISTER FROM LOT 3160517 TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT, DURING PATIENT USE, THE CO2 LEVEL WENT UP. THERE WAS NO REPORTED PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
591631 CARESTATION 650 ANESTHESIA GAS MACHINE BSZ GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)

Patients

Seq Age Sex Outcome Treatment