1823260-2017-01788
Report
- Report Number
- 1823260-2017-01788
- Event Type
- Malfunction
- Date Received
- August 22, 2017
- Date of Event
- August 1, 2017
- Report Date
- August 22, 2017
- PMA / PMN Number
- K100853
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(6). (B)(4).
THE CUSTOMER STATED THAT THEY NOTICED QUESTIONABLE RESULTS FOR SOME PATIENT SAMPLES TESTED FOR ALB2 ALBUMIN GEN.2 (ALB) ON THE COBAS 8000 C (701) MODULE - C701. THIS PROMPTED THEM TO REPEAT 1000 PATIENT SAMPLES. THE CUSTOMER PROVIDED DATA FOR TWO PATIENT SAMPLES THAT WERE QUESTIONED. OF THESE, ONE HAD AN ERRONEOUS RESULT. THE ERRONEOUS RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER ALSO STATED THAT THEY WERE SEEING ERRATIC BEHAVIOR FOR THE ENZYMATIC CREATININE TEST RUN ON THE OTHER DISK OF THE ANALYZER. NO FURTHER DETAILS WERE PROVIDED REGARDING THIS. THE SAMPLE INITIALLY RESULTED WITH AN ALB VALUE OF 58.6 G/L. THE SAMPLE WAS REPEATED ON (B)(6) 2017, RESULTING AS 39.3 G/L. THE SAMPLE WAS ALSO REPEATED ON (B)(6) 2017, RESULTING AS 40.8 G/L. NO ADVERSE EVENTS WERE ALLEGED TO HAVE OCCURRED WITH THE PATIENT. THE CUSTOMER RAN PRECISION STUDIES. THE CUSTOMER CHANGED ALL REAGENT AND SAMPLE PROBES. THESE ACTIONS DID NOT RESOLVE THE ISSUE. THE FIELD SERVICE ENGINEER CHECKED WASHING ON THE ANALYZER AND THIS WAS OK. HE CHECKED GEAR PUMP PRESSURE AND PROBE WASHING AND THESE WERE OK. HE CHECKED THE HIGH CONCENTRATION VACUUM AND PRESSURE WAS A BIT LOW. HE REPLACED A VALVE AND DAMAGED TUBING. CALIBRATION AND QUALITY CONTROLS WERE OK. THE ISSUE WAS DUE TO AGING OF THE REPLACED VALVE.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |