STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2017-01194
- Event Type
- Injury
- Date Received
- August 21, 2017
- Date of Event
- July 20, 2017
- Report Date
- September 18, 2017
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON AUG 18, 2017 WITH LOT NUMBER 1230512. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: FOLD CREASES, WEAR ABRASION. A LEAK TEST OF THE RETURNED DEVICE WAS PERFORMED AND FOUND TWO OPENINGS. A MICROSCOPIC ANALYSIS OF THE RETURNED DEVICE IDENTIFIED TWO CURVED STRIATED OPENING ON ANTERIOR AND RADIUS SIDE ASSESSED AS SURGICAL DAMAGE, PARTIAL DELAMINATION OF DIAPHERM FLANGE DISK ASSESSED AS ADHESIVE FAILURE. FILL INSPECTION WAS PERFORMED AND IDENTIFIED NO BLOCKAGE IN THE VALVE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: TWO CURVED STRIATED OPENINGS ASSESSED AS SURGICAL DAMAGE.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. DEVICE LABELING: POTENTIAL ADVERSE EVENTS THAT MAY OCCUR WITH SALINE-FILLED BREAST IMPLANT SURGERY INCLUDE: REOPERATION, PAIN, WRINKLING, ASYMMETRY, IMPLANT PALPABILITY/VISIBILITY, IMPLANT REMOVAL, CAPSULAR CONTRACTURE, CHANGES IN NIPPLE AND BREAST SENSATION, IMPLANT DISPLACEMENT/MIGRATION, IMPLANT DEFLATION, SCARRING, INFECTION, HEMATOMA/SEROMA, BREASTFEEDING COMPLICATIONS, IMPLANT EXTRUSION, NECROSIS, DELAYED WOUND HEALING, BREAST TISSUE ATROPHY/CHEST WALL DEFORMITY, CALCIUM DEPOSITS, AND LYMPHADENOPATHY. DEFLATION ¿ BREAST IMPLANTS ARE NOT LIFETIME DEVICES. SALINE BREAST IMPLANTS DEFLATE WHEN THE SHELL DEVELOPS A TEAR OR HOLE. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT THEY ARE MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING THINGS MAY CAUSE IMPLANTS TO DEFLATE: DAMAGE BY SURGICAL INSTRUMENTS; FOLDING OR WRINKLING OF THE IMPLANT SHELL; EXCESSIVE FORCE TO THE CHEST (E.G., DURING CLOSED CAPSULOTOMY, WHICH IS CONTRAINDICATED); TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. LABORATORY STUDIES HAVE IDENTIFIED SOME OF THE CAUSES OF DEFLATION FOR ALLERGAN¿S PRODUCT; HOWEVER, IT IS NOT CONCLUSIVELY KNOWN WHETHER THESE TESTS HAVE IDENTIFIED ALL CAUSES OF DEFLATION.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE WAS EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 588712 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 1230512 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Required Intervention |