STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2017-01272
- Event Type
- Injury
- Date Received
- August 21, 2017
- Date of Event
- May 27, 2017
- Report Date
- August 21, 2017
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. DEVICE ANALYSIS RESULTS: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON (B)(6) 2017 WITH LOT NUMBER 2042797. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED AN OPENING, A CREASE FOLD, YELLOW PARTICLES, AND WEAR ABRASION. A LEAK TEST WAS PERFORMED AND FOUND AN OPENING ON POSTERIOR SIDE. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED A SHARP EDGE OPENING ASSESSED AS UNIDENTIFIED TEAR OPENING. THE THICKNESS WAS FOUND TO BE WITHIN SPECIFICATION. A FILL TEST INSPECTION WAS PERFORMED AND FOUND NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP OPENING ON POSTERIOR DUE TO AN UNIDENTIFIED (TEAR) OPENING. DEVICE LABELING ADDRESSES: ¿POTENTIAL ADVERSE EVENTS THAT MAY OCCUR WITH SALINE-FILLED BREAST IMPLANT SURGERY INCLUDE: REOPERATION, PAIN, WRINKLING, ASYMMETRY, IMPLANT PALPABILITY/VISIBILITY, IMPLANT REMOVAL, CAPSULAR CONTRACTURE, CHANGES IN NIPPLE AND BREAST SENSATION, IMPLANT DISPLACEMENT/MIGRATION, IMPLANT DEFLATION, SCARRING, INFECTION, HEMATOMA/SEROMA, BREASTFEEDING COMPLICATIONS, IMPLANT EXTRUSION, NECROSIS, DELAYED WOUND HEALING, BREAST TISSUE ATROPHY/CHEST WALL DEFORMITY, CALCIUM DEPOSITS, AND LYMPHADENOPATHY. DEFLATION ¿ BREAST IMPLANTS ARE NOT LIFETIME DEVICES. SALINE BREAST IMPLANTS DEFLATE WHEN THE SHELL DEVELOPS A TEAR OR HOLE. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT THEY ARE MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING THINGS MAY CAUSE IMPLANTS TO DEFLATE: DAMAGE BY SURGICAL INSTRUMENTS; FOLDING OR WRINKLING OF THE IMPLANT SHELL; EXCESSIVE FORCE TO THE CHEST (E.G., DURING CLOSED CAPSULOTOMY, WHICH IS CONTRAINDICATED); TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. LABORATORY STUDIES HAVE IDENTIFIED SOME OF THE CAUSES OF DEFLATION FOR ALLERGAN¿S PRODUCT; HOWEVER, IT IS NOT CONCLUSIVELY KNOWN WHETHER THESE TESTS HAVE IDENTIFIED ALL CAUSES OF DEFLATION.¿
HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. THE DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589083 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 2042797 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |