130 DEG AIMING ARM F/TROCHANTERIC FIXATION NAILS
Report
- Report Number
- 2530088-2017-10253
- Event Type
- Malfunction
- Date Received
- August 18, 2017
- Date of Event
- July 23, 2017
- Report Date
- July 25, 2017
- Manufacturer
- SYNTHES BRANDYWINE
- Product Code
- FZX
- UDI-DI
- 10886982196019
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
PATIENT INFORMATION NOT AVAILABLE FOR REPORTING. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. DHR REVIEW FOR PART: # 357.366, SYNTHES LOT # 5223477. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURING OF THIS PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. RELEASE TO WAREHOUSE DATE: APRIL 18, 2006. (B)(4). DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
CUSTOMER QUALITY COMPLETED AN INVESTIGATION OF THE RETURNED DEVICES. THE FOLLOWING DEVICES WERE RECEIVED 357.411, LOT 4544026 AND 357.366, LOT 5223477 WITH THE FOLLOWING COMPLAINT DESCRIPTION,"IT WAS REPORTED THAT DURING A TFN (TROCHANTERIC FIXATION NAIL) SURGERY, THE 130 DEGREE AIMING ARM AND THE INSERTION HANDLE DID NOT LINE UP PROPERLY WHEN INSERTING THE DISTAL LOCKING SCREW. THERE ARE TWO DEVICES INVOLVED IN THIS COMPLAINT. A DEVICE HISTORY REVIEW (DHR), DEVICE INSPECTION, AND FUNCTIONAL TEST REVIEW WERE PERFORMED AS PART OF THIS INVESTIGATION. THE COMPLAINT IS UNCONFIRMED. CLINICALLY, ALIGNMENT OF THE LOCKING SCREW WAS ACCOMPLISHED WHEN THE USER RE-TIGHTENED THE HANDLE AND AIMING ARM AND DRIVE CAP. FUNCTIONAL REVIEW: REPLICATION CANNOT BE PERFORMED DUE TO NOT HAVING FULL DETAILS ABOUT THE CASE AND NOT HAVING ALL MATING PARTS. UPON INSPECTION, THE AIMING ARM FASTENED TO THE INSERTION HANDLE AS INDICATED. THE COMPLAINT IS UNCONFIRMED. CLINICALLY, ALIGNMENT OF THE LOCKING SCREW WAS ACCOMPLISHED WHEN THE USER RE-TIGHTENED THE HANDLE AND AIMING ARM AND DRIVE CAP. NO NEW MALFUNCTIONS WERE IDENTIFIED AS A RESULT OF THIS INVESTIGATION. THE 357.411, LOT 4544026, INSERTION HANDLE AND 357.366, LOT 5223477, AIMING ARM IS USED IN CONJUNCTION WITH EACH OTHER ALONG WITH THE TROCANTERIC NAILS FOR BLADE ALIGNMENT, PER TFN ADVANCED PROXIMAL FEMORAL NAILING SYSTEM TECHNIQUE GUIDE. VISUAL INSPECTION: UPON INSPECTION, THE AIMING ARM FASTENED TO THE INSERTION HANDLE AS INDICATED. THE DEVICES SHOWED WEAR OFTEN ASSOCIATED WITH REPEAT USE AFTER 11 YEARS OF FIELD USE. THE COMPLAINT IS UNCONFIRMED. CLINICALLY, ALIGNMENT OF THE LOCKING SCREW WAS ACCOMPLISHED WHEN THE USER RE-TIGHTENED THE HANDLE AND AIMING ARM AND DRIVE CAP. DHR REVIEW FOR PART: PART: #357.366 -SYNTHES LOT # 5223477 NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURING OF THIS PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DRAWING REVIEW: DRAWINGS WERE REVIEWED AND THE HOLE PATTERNS VERIFIED THE PARTS ARE ACCEPTABLE AND MEET PRINT. DCRM REVIEW: A DCRM REVIEW AND/OR OCCURRENCE RATE CALCULATION MUST BE PERFORMED FOR ALL COMPLAINT PARTS WHICH RESULTED IN A CLASSIFICATION OF SERIOUS INJURY THAT WERE NOT GENERATED FROM A LITERATURE ARTICLE. THEREFORE NO DCRM CALCULATION WILL BE PERFORMED FOR THIS COMPLAINT. NO DEFINITE ROOT CAUSE CAN BE DETERMINED WHY THE ALIGNMENT TO THE LOCKING SCREW OF THE INSERTION HANDLE AND AIMING GUIDE DID NOT TAKE PLACE. UPON INSPECTION, THE AIMING ARM FASTENED TO THE INSERTION HANDLE AS INDICATED. THE COMPLAINT IS UNCONFIRMED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING A TROCHANTERIC FIXATION NAIL (TFN) SURGERY ON (B)(6) 2017, THE 130 DEGREE AIMING ARM AND THE INSERTION HANDLE DID NOT LINE UP PROPERLY WHEN INSERTING THE DISTAL LOCKING SCREW. THE NAIL AND THE HELICAL BLADE WERE BOTH ALREADY IN PLACE. THE SURGEON ADJUSTED AND TIGHTENED THE AIMING ARM, HANDLE AND THE DRIVING CAP, AND WAS THEN ABLE TO USE THE INSTRUMENTS TO IMPLANT THE DISTAL SCREW AS INTENDED. THERE WERE NO FRAGMENTS GENERATED, NO SURGICAL DELAY, AND NO ADDITIONAL MEDICAL INTERVENTION. THE SURGERY WAS COMPLETED, AND THE PATIENT OUTCOME WAS STABLE. CONCOMITANT DEVICES REPORTED: TFN NAIL (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1); HELICAL BLADE (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1); DISTAL LOCKING SCREW (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1); DRIVING CAP (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) 130 DEGREE AIMING ARM (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 584623 | 130 DEG AIMING ARM F/TROCHANTERIC FIXATION NAILS | GUIDE | FZX | SYNTHES BRANDYWINE | 5223477 | 10886982196019 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |