FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 6796507 · Received August 15, 2017

Report

Report Number
9617229-2017-00996
Event Type
Injury
Date Received
August 15, 2017
Date of Event
July 6, 2017
Report Date
September 29, 2017
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. DEVICE LABELING: POTENTIAL ADVERSE EVENTS THAT MAY OCCUR WITH SALINE-FILLED BREAST IMPLANT SURGERY INCLUDE: REOPERATION, PAIN, WRINKLING, ASYMMETRY, IMPLANT PALPABILITY/VISIBILITY, IMPLANT REMOVAL, CAPSULAR CONTRACTURE, CHANGES IN NIPPLE AND BREAST SENSATION, IMPLANT DISPLACEMENT/MIGRATION, IMPLANT DEFLATION, SCARRING, INFECTION, HEMATOMA/SEROMA, BREASTFEEDING COMPLICATIONS, IMPLANT EXTRUSION, NECROSIS, DELAYED WOUND HEALING, BREAST TISSUE ATROPHY/CHEST WALL DEFORMITY, CALCIUM DEPOSITS, AND LYMPHADENOPATHY. DEFLATION ¿ BREAST IMPLANTS ARE NOT LIFETIME DEVICES. SALINE BREAST IMPLANTS DEFLATE WHEN THE SHELL DEVELOPS A TEAR OR HOLE. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT THEY ARE MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING THINGS MAY CAUSE IMPLANTS TO DEFLATE: DAMAGE BY SURGICAL INSTRUMENTS; FOLDING OR WRINKLING OF THE IMPLANT SHELL; EXCESSIVE FORCE TO THE CHEST (E.G., DURING CLOSED CAPSULOTOMY, WHICH IS CONTRAINDICATED); TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. LABORATORY STUDIES HAVE IDENTIFIED SOME OF THE CAUSES OF DEFLATION FOR ALLERGAN¿S PRODUCT; HOWEVER, IT IS NOT CONCLUSIVELY KNOWN WHETHER THESE TESTS HAVE IDENTIFIED ALL CAUSES OF DEFLATION.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON SEPTEMBER 05, 2017 WITH LOT NUMBER 2191639. LEAK TEST WAS PERFORMED AND FOUND OPENING ON POSTERIOR AND ANTERIOR SIDE VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: CREASE FOLD, WEAR ABRASION AND OPENING. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH ONE CREASE SHARP IN THE ANTERIOR SIDE AND ONE OPENING STRIATED IN THE PATCH OVERALL. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A CREASE SHARP IN THE ANTERIOR SIDE. A STRIATED EDGE OPENING ON PATCH OVERALL SIDE DUE TO SURGICAL DAMAGE.

Description of Event or Problem · 1

PATIENT REPORTED A RIGHT SIDE DEFLATION. HEALTHCARE PROFESSIONAL CONFIRMED THE EVENT. THE DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 1

THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574252 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 2191639

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention