M2A 38MM MOD HD+9MM NK NO SKRT
Report
- Report Number
- 0001825034-2017-06372
- Event Type
- Injury
- Date Received
- August 15, 2017
- Date of Event
- September 25, 2012
- Report Date
- February 21, 2018
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWA
- PMA / PMN Number
- PK0111110
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(6). CONCOMITANT PRODUCTS - PART: X11-180308 NAME: BI-METRIC/X POR NC LAT LOT:144910. THE INVESTIGATION IS STILL IN PROCESS. ONCE THE INVESTIGATION IS COMPLETE A FOLLOW UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORTS: 0001825034-2017-06371.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. 15-105054 M2A 1 PC SHELL 38MMX54MM 043790, X11-180308 BI-METRIC/X POR NC LAT 8X120 144910. REPORTED EVENT WAS CONFIRMED BY REVIEW OF THE PROVIDED OP NOTES, WHICH NOTED A MINOR AMOUNT OF SYNOVIAL IRRITATION, AND A LARGE PROJECTION OF BONE ANTERIOR ASPECT OF THE ACETABULAR, WHICH WAS BELIEVED TO BE A SIGNIFICANT CAUSE OF PAIN. DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE COULD NOT BE DETERMINED WITH INFORMATION AVAILABLE. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT THIS TYPE OF EVENT CAN OCCUR AND RISKS ARE ADDRESSED IN THE ASSOCIATED RISK DOCUMENTATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT A PATIENT WAS REVISED DUE TO GROIN PAIN, ELEVATED METAL ION LEVELS, AND ADVERSE REACTION TO METAL DEBRIS. IT WAS NOTED THAT THERE WERE SOME SYNOVIAL IRRITATION UPON ENTERING THE JOINT CAPSULE. THERE WAS ALSO A LARGE BONE PROJECTION OVER THE ANTERIOR ASPECT OF THE CUP. WHILE ATTEMPTING TO REMOVE THE OSETEOME A PORTION OF THE ANTERIOR SHELL CAME WITH IT. THEREFORE, THE CUP WAS REMOVED. NO COMPLICATIONS INDICATED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
IT WAS REPORTED THE PATIENT WAS INDICATED FOR A RIGHT HIP REVISION PROCEDURE DUE TO PERSISTENT GROIN PAIN AS WELL AS POSSIBLE ADVERSE TISSUE REACTION DUE TO ELEVATED METAL ION LEVELS. THE FEMORAL HEAD AND ACETABULAR COMPONENT WERE REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 574745 | M2A 38MM MOD HD+9MM NK NO SKRT | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 568320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Hospitalization| R |