FDA Adverse Event Injury Summary report: N

HEAL COLLAR 3.5X4.5, 5MM

MDR report key: 6791300 · Received August 14, 2017

Report

Report Number
0001038806-2017-00492
Event Type
Injury
Date Received
August 14, 2017
Date of Event
May 17, 2017
Report Date
October 18, 2017
Manufacturer
ZIMMER DENTAL
Product Code
NHA
PMA / PMN Number
PK111852
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE LOT # WAS NOT PROVIDED/UNKNOWN.

Additional Manufacturer Narrative · 1

ONE ZIMMER DENTAL HEAL COLLAR WAS RETURNED FOR INSPECTION. VISUAL INSPECTION REVEALED WEAR ABOUT THE COLLAR AND THE THREADS. THE HEX HEAD WAS ALSO SLIGHTLY WORN WITH TISSUE ATTACHMENT, BUT FUNCTIONAL. PRODUCT DIMENSIONS WERE FOUND TO CONFORM TO PRINT SPECIFICATIONS WHERE MEASURED AGAINST DRAWING REF ECO# 112456 REV A. THE IMPLANT THAT ALLEGEDLY WOULD NOT MATE WAS NOT RETURNED FOR INSPECTION; THEREFORE THE REPORTED EVENT COULD NOT BE VERIFIED. NO DEVICE LOT NUMBER WAS PROVIDED SO A DEVICE HISTORY RECORD REVIEW AND A COMPLAINT HISTORY REVIEW COULD NOT BE PERFORMED. APPROPRIATE DOCUMENTATION WAS REVIEWED AND THE FOLLOWING INFORMATION WAS IDENTIFIED: INSTRUCTIONS FOR USE FOR HEALING COLLARS, HEALING SCREWS, SURGICAL COVER SCREWS AND TEMPORARY GINGIVAL CUFFS¿ 9658 REV 1-01/15. HEALING COLLARS ¿ FOR USE WITH TAPERED SCREW-VENT®, SCREW-VENT® AND TRABECULAR METAL¿ IMPLANTS SELECT THE APPROPRIATE SIZE HEALING COLLAR FOR THE PLATFORM OF THE IMPLANT AND WITH APPROPRIATE HEIGHT AND EMERGENCE PROFILE TO FIT THE EXISTING OR DESIRED TISSUE CONTOUR OF THE SITE. THE HEALING COLLARS ARE ANODIZED WITH COLOR-CODING ON THE LOWER PORTION TO INDICATE THE IMPLANT PLATFORM DIAMETER (FIGURE A). THE TOP SURFACE OF THE HEALING COLLAR IS ETCHED WITH THREE NUMBERS (FIGURE B) TO REFERENCE IMPLANT PLATFORM DIAMETER (LEFT), EMERGENCE PROFILE DIAMETER (TOP RIGHT) AND CUFF HEIGHT (LOWER RIGHT). THE NUMBERS FOR IMPLANT PLATFORM AND EMERGENCE PROFILE SHOW ONLY THE INITIAL DIGIT OF THE RELATED MEASUREMENT. PROBABLE CAUSES FOR THE COMPLAINT COULD NOT BE DETERMINED WITH THE INFORMATION PROVIDED AS THE EVENT CANNOT BE RECREATED. COMPLAINT IS THEREFORE NON-VERIFIABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT HEALING COLLAR DOES NOT ASSEMBLE IN TOOTH LOCATION#30. PATIENT HAS BEEN RESCHEDULED FOR NEW PROCEDURE FOR HEALING COLLAR PLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
572361 HEAL COLLAR 3.5X4.5, 5MM HEALING ABUTMENT NHA ZIMMER DENTAL

Patients

Seq Age Sex Outcome Treatment
1 26 YR Required Intervention