5.0MM TI LOCKING SCREW W/T25 STARDRIVE 64MM F/IM NAIL-STER
Report
- Report Number
- 1719045-2017-10773
- Event Type
- Injury
- Date Received
- August 11, 2017
- Report Date
- July 27, 2017
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- HWC
- PMA / PMN Number
- K000089
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
(B)(6). (B)(4). COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. DHR REVIEW WAS CONDUCTED. MANUFACTURING LOCATION: (B)(4). MANUFACTURING DATE: 01-JUL-2015 EXPIRATION DATE: 31-MAY-2024 PART #: 04.005.554S, LOT#: 9843590 (STERILE) -. QUANTITY (B)(4). COMPONENT PARTS REVIEWED: PART 04.005M554, BP57 LOT: 9804571. 5.0MM TI LOCKING SCREW W/T25. LOT WAS RELEASED TO BP57 ON 13-MAY-2015. COMPONENT PART 04.005.554.999, 5.0MM TI SCREW BLANK 64MM, BP55 LOT: 7883612. INSPECTION SHEET FOR WHIRL THREAD, VIBE, FLUTE FINAL INSPECTION (B)(4) REV: H MET INSPECTION ACCEPTANCE CRITERIA. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. ¿STERILITY DOCUMENTATION WAS REVIEWED AND DETERMINED TO BE CONFORMING.¿ DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORT THAT THE PATIENT HAD ORIGINAL SURGERY ON (B)(6) 2016 FOR TREATMENT OF A FEMUR FRACTURE WERE THE SURGEON USED THE (TFNA) TROCHANTERIC FIXATION NAIL ADVANCED IMPLANTS. PATIENT WAS IMPLANTED WITH ONE (1) TFNA 12MM/130 DEG TI CANNULATED NAIL, ONE (1) LAG SCREW AND THREE (3) DISTAL LOCKING SCREWS. ON AN UNKNOWN DATE, POST-OPERATIVE, PATIENT PRESENTED WITH A NON-UNION AT THE INTERTROCHANTERIC DISTAL THIRD PORTION OF THE FEMUR BONE. ON (B)(6) 2017, SURGEON REMOVED ALL IMPLANTS AND REVISED THE PATIENT TO A COMPETITOR¿S NATURAL NAIL. IT WAS REPORTED THAT NO FRAGMENTS WERE GENERATED DURING IMPLANT REMOVAL. REVISION SURGERY WAS COMPLETED SUCCESSFULLY WITH NO TIME DELAY. PATIENT IS REPORTED IN STABLE CONDITION. THIS REPORT IS FOR ONE (1) 5.0MM LOCKING SCREW FOR IM NAIL THIS IS REPORT 5 OF 5 FOR (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 568969 | 5.0MM TI LOCKING SCREW W/T25 STARDRIVE 64MM F/IM NAIL-STER | SCREW,FIXATION,BONE | HWC | SYNTHES MONUMENT | 9843590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |