HEAL COLLAR 3.5X4.5, 3MM
Report
- Report Number
- 0001038806-2017-00488
- Event Type
- Malfunction
- Date Received
- August 10, 2017
- Date of Event
- June 29, 2017
- Report Date
- October 18, 2017
- Manufacturer
- ZIMMER DENTAL
- Product Code
- NHA
- PMA / PMN Number
- PK111852
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PATIENT INFORMATION IS UNKNOWN/NOT PROVIDED.
ONE ZIMMER DENTAL HEAL COLLAR WAS RETURNED FOR INSPECTION. VISUAL INSPECTION REVEALED WEAR ABOUT THE COLLAR AND THE THREADS. THE HEX HEAD WAS ALSO SLIGHTLY WORN WITH TISSUE ATTACHMENT, BUT FUNCTIONAL. PRODUCT DIMENSIONS WERE FOUND TO CONFORM TO PRINT SPECIFICATIONS WHERE MEASURED AGAINST DRAWING REF ECO# 112456 REV A. THE IMPLANT THAT ALLEGEDLY WOULD NOT MATE WAS NOT RETURNED FOR INSPECTION, THEREFORE THE REPORTED EVENT COULD NOT BE VERIFIED. A DEVICE HISTORY REVIEW WAS PERFORMED AND NO RELATED NONCONFORMANCE¿S WERE NOTED. A COMPLAINT HISTORY SEARCH WAS PERFORMED USING OUR COMPLAINT HANDLING SYSTEM AND THERE WERE NO ADDITIONAL COMPLAINTS GENERATED AGAINST THIS LOT NUMBER. APPROPRIATE DOCUMENTATION WAS REVIEWED AND THE FOLLOWING INFORMATION WAS IDENTIFIED: INSTRUCTIONS FOR USE FOR HEALING COLLARS, HEALING SCREWS, SURGICAL COVER SCREWS AND TEMPORARY GINGIVAL CUFFS¿ 9658 REV 1-01/15. HEALING COLLARS ¿ FOR USE WITH TAPERED SCREW-VENT®, SCREW-VENT® AND TRABECULAR METAL¿ IMPLANTS SELECT THE APPROPRIATE SIZE HEALING COLLAR FOR THE PLATFORM OF THE IMPLANT AND WITH APPROPRIATE HEIGHT AND EMERGENCE PROFILE TO FIT THE EXISTING OR DESIRED TISSUE CONTOUR OF THE SITE. THE HEALING COLLARS ARE ANODIZED WITH COLOR-CODING ON THE LOWER PORTION TO INDICATE THE IMPLANT PLATFORM DIAMETER (FIGURE A). THE TOP SURFACE OF THE HEALING COLLAR IS ETCHED WITH THREE NUMBERS (FIGURE B) TO REFERENCE IMPLANT PLATFORM DIAMETER (LEFT), EMERGENCE PROFILE DIAMETER (TOP RIGHT) AND CUFF HEIGHT (LOWER RIGHT). THE NUMBERS FOR IMPLANT A ROOT CAUSE FOR THE COMPLAINT COULD NOT BE DETERMINED WITH THE INFORMATION PROVIDED AS THE EVENT CANNOT BE RECREATED. COMPLAINT IS THEREFORE NON-VERIFIABLE. UDI NUMBER: (B)(4).
THE DOCTOR REPORTED THAT DURING A PROCEDURE ON (B)(6) 2017, THE HEALING COLLAR (HC343) DID NOT SCREW INTO THE IMPLANT. DOCTOR TRIED TO SCREW THE HEALING COLLAR INTO SEVERAL ANALOGS BUT UNSUCCESSFULLY. DOCTOR COMPLETED THE PROCEDURE BY PLACING ANOTHER HEALING COLLAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 564735 | HEAL COLLAR 3.5X4.5, 3MM | HEALING ABUTMENT | NHA | ZIMMER DENTAL | 63346429 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |