PHOENIX 3.5MM INSERTER CONNECTOR SHORT
Report
- Report Number
- 0001825034-2017-05891
- Event Type
- Injury
- Date Received
- August 8, 2017
- Date of Event
- July 5, 2017
- Report Date
- May 23, 2018
- Manufacturer
- BIOMET TRAUMA
- Product Code
- LXH
- PMA / PMN Number
- PEXEMPT
- Removal / Correction Number
- Z-2017-00011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
PT IDENTIFIER: - (B)(4). CONCOMITANT MEDICAL PRODUCT - OFFSET END CAP 11X5MM, PART #: 14-441281, LOT #279510. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2017-05891, AND 0001825034-2017-05892.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. INITIAL REPORT REPORTED LOT#: ZB151101. HOWEVER UPON FURTHER REVIEW IT IS FOUND THAT LOT ZB151101 WAS ISSUED TO 2 LOT #'S 335080 AND 335090. AS IT IS UNKNOWN WHICH LOT THIS COMPLAINT PERTAINS TO WE WILL LIST BOTH. LOT#: 335080, MANUFACTURE: DEC 9, 2015. LOT#: 335090, MANUFACTURE: DEC 8, 2015. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. CONCOMITANT MEDICAL PRODUCT- 3.5MM INSERTER SHORT CATALOG#: 14-441046 LOT#:098710, 3.5MM INSERTER SHORT CATALOG#: 14-441046 LOT#:615580, 3.5MM INSERTER SHORT CATALOG#: 14-441046 LOT#:843800, UNKNOWN END CAP 10MM CATALOG#:UNK LOT#:UNK, OFFSET END CAP 11X5MM CATALOG#:14-441281 LOT#:311710. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. THE PRODUCT WAS RETURNED FOR EVALUATION. THE RETURNED TIP OF THE CONNECTOR IS BROKEN INTO THE END CAP. THE INSERTER TIP IS ALSO FRACTURED. PARTS ARE NOTED TO BE VISUALLY CONFORMING TO PRINT SPECIFICATIONS OTHERWISE. DUE TO THE MISSING PIECE OF THE CONNECTOR AND THE FRACTURED PART OF THE CONNECTOR, ACCURATE FUNCTIONAL AND DIMENSIONAL ANALYSIS CANNOT BE PERFORMED. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE OF THE CONNECTOR BREAKING INTO THE END CAP IS RELATED TO DESIGN INPUT AND VERIFICATION DEFICIENCY. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT DURING A TIBIA FRACTURE NAILING PROCEDURE, THAT BOTH THE INSERTER CONNECTOR AND INSERTER FRACTURED, WHICH LED TO THE REMOVAL OF THE END CAP ALREADY IMPLANTED IN THE PATIENT, AS THE INSERTER CONNECTOR REMAINED IN THE ORIGINAL END CAP. A 10MM END CAP WAS IMPLANTED WITH MINIMAL DELAY TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 554865 | PHOENIX 3.5MM INSERTER CONNECTOR SHORT | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | BIOMET TRAUMA | N/A | REFERENCE H10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |