MDR report key: 67723 · Received February 7, 1997

Report

Report Number
2027242-1997-90040
Date Received
February 7, 1997
Report Date
January 10, 1997
Manufacturer
DIAGNOSTIC SOLUTIONS, INC.
Product Code
CFR
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

A PHARMACY CLERK REPORTED THAT A CUSTOMER HAD RETURNED A BOX OF TEST STRIPS COMPLAINING OF INACCURATELY HIGH READINGS. THE CUSTOMER TOLD THE PHARMACY CLERK THAT HE HAD PERFORMED APPROX SIX BLOOD GLUCOSE TESTS AND THAT EACH TIME HE PERFORMED COMPARISON TESTS WITH ANOTHER TEST STRIPS AND OBTAINED RESULTS OF 123, 120, AND 180 MG/DL RESPECTIVELY. THE CUSTOMER PERFORMED A CONTROL SOLUTION STATED AND OBTAINED A RESULT OF 170 VERSUS A NORMAL CONTROL SOLUTION RANGE OF 106-158 MG/DL. THE CUSTOMER STATEED THAT THE CODE WAS SET CORRECTLY FOR ALL TESTS. HE ALS PERFORMED A CHECK STRIP TEST AND OBTAINED A RESULT THAT WAS WITHIN RANGE (SPECIFIC RESULT UNKNOWN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK CHECK ONE BLOOD GLUCOSE TEST STRIPS FOR USE WITH ONE TOUCH METERS CFR DIAGNOSTIC SOLUTIONS, INC. * 16199A

Patients

Seq Age Sex Outcome Treatment
1 *