1823260-2017-01624
Report
- Report Number
- 1823260-2017-01624
- Event Type
- Malfunction
- Date Received
- August 7, 2017
- Date of Event
- July 13, 2017
- Report Date
- August 13, 2017
- PMA / PMN Number
- K100853
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
AGE AT TIME OF THE EVENT WAS UPDATED. A REVIEW OF THE ALARM TRACE SHOWS SEVERAL "ABNORMAL SAMPLE ASPIRATION" MESSAGES. A SPECIFIC ROOT CAUSE WAS NOT IDENTIFIED. A POSSIBLE ROOT CAUSE IS THAT THERE WAS FOAM ON THE SAMPLE.
AGE AT TIME OF EVENT ASKU: THE PATIENT IS DOCUMENTED AS BEING (B)(6) AND (B)(6) OLD. CLARIFICATION OF THE CORRECT AGE HAS BEEN REQUESTED BUT HAS NOT BEEN PROVIDED. THIS EVENT OCCURRED IN (B)(6). (B)(4).
THE CUSTOMER COMPLAINED OF AN ERRONEOUS LOW RESULT FOR 1 PATIENT TESTED FOR ELECSYS HCG + BETA TEST SYSTEM (HCG + B) ON A COBAS 8000 E 602 MODULE. THE INITIAL HCG + B RESULT WAS 0.277 MIU/ML. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE WAS REPEATED AND THE RESULT WAS 258.2 MIU/ML. THE REPEAT RESULT WAS CONSIDERED TO BE CORRECT. THERE WAS NO ALLEGATION THAT AN ADVERSE EVENT OCCURRED. THE HCG + B REAGENT LOT NUMBER WAS 210151. THE EXPIRATION DATE WAS NOT PROVIDED. THE QUALITY CONTROL (QC) RESULTS WERE ACCEPTABLE. THE PINCH TUBE WAS LAST REPLACED ON (B)(6) 2017. LIQUID FLOW CLEANING IS PERFORMED EVERY 3 DAYS. INSTRUMENT PERFORMANCE TEST RESULTS WERE ACCEPTABLE. THE POSSIBLE ROOT CAUSE OF THE ISSUE MAY HAVE BEEN FOAM ON THE SAMPLE AS A BUBBLE WAS DETECTED ON THE SURFACE OF THE SAMPLE.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |