FDA Adverse Event Malfunction Summary report: N

VASOVIEW 7 XB

MDR report key: 6768913 · Received August 4, 2017

Report

Report Number
2242352-2017-00748
Event Type
Malfunction
Date Received
August 4, 2017
Date of Event
July 12, 2017
Report Date
August 4, 2017
Manufacturer
MAQUET CV
Product Code
GEI
PMA / PMN Number
K041981
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE VV7 BISECTOR DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. SIGNS OF CLINICAL USE WAS OBSERVED NO EVIDENCE OF BLOOD WERE OBSERVED. NO VISUAL DEFECTS WERE OBSERVED. THE DEVICE WAS EVALUATED FOR ELECTRICAL FUNCTION. A PRE-CAUTERY TEST WAS PERFORMED AND REPEATED 10 TIMES OVER A 10 MINUTE PERIOD ACCORDING TO THE PROCEDURE IN THE INSTRUCTIONS FOR USE. THE DEVICE PASSED THE PRE-CAUTERY TEST WITH A REFERENCE GENERATOR (RECOMMENDED SETTING OF 18) AND REFERENCE BIPOLAR CORD. THE BIPOLAR CORD CONNECTION WAS MANIPULATED DURING ACTIVATION. THE DEVICE REMAINED ACTIVE AND NO INTERMITTENT CONTINUITY WAS OBSERVED. THE DEVICE PRODUCED STEAM AND THE SALINE WAS OBSERVED TO ¿BOIL¿ ON THE TEST GAUZE EACH TIME. BASED ON THE RESULTS OF THE EVALUATION, THE REPORTED COMPLAINT WAS NOT CONFIRMED FOR THE REPORTED FAILURE MODE "FAILURE TO DELIVER ENERGY".

Additional Manufacturer Narrative · 1

(B)(4). A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. THE DEVICE HAS BEEN RETURNED TO THE FACTORY AND IS BEING EVALUATED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIW 7XB CUTTING DEVICE DID NOT WORK. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIW 7XB CUTTING DEVICE DID NOT WORK. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
549487 VASOVIEW 7 XB ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI MAQUET CV 25132654

Patients

Seq Age Sex Outcome Treatment
1