VASOVIEW 7 XB
Report
- Report Number
- 2242352-2017-00748
- Event Type
- Malfunction
- Date Received
- August 4, 2017
- Date of Event
- July 12, 2017
- Report Date
- August 4, 2017
- Manufacturer
- MAQUET CV
- Product Code
- GEI
- PMA / PMN Number
- K041981
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE VV7 BISECTOR DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. SIGNS OF CLINICAL USE WAS OBSERVED NO EVIDENCE OF BLOOD WERE OBSERVED. NO VISUAL DEFECTS WERE OBSERVED. THE DEVICE WAS EVALUATED FOR ELECTRICAL FUNCTION. A PRE-CAUTERY TEST WAS PERFORMED AND REPEATED 10 TIMES OVER A 10 MINUTE PERIOD ACCORDING TO THE PROCEDURE IN THE INSTRUCTIONS FOR USE. THE DEVICE PASSED THE PRE-CAUTERY TEST WITH A REFERENCE GENERATOR (RECOMMENDED SETTING OF 18) AND REFERENCE BIPOLAR CORD. THE BIPOLAR CORD CONNECTION WAS MANIPULATED DURING ACTIVATION. THE DEVICE REMAINED ACTIVE AND NO INTERMITTENT CONTINUITY WAS OBSERVED. THE DEVICE PRODUCED STEAM AND THE SALINE WAS OBSERVED TO ¿BOIL¿ ON THE TEST GAUZE EACH TIME. BASED ON THE RESULTS OF THE EVALUATION, THE REPORTED COMPLAINT WAS NOT CONFIRMED FOR THE REPORTED FAILURE MODE "FAILURE TO DELIVER ENERGY".
(B)(4). A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. THE DEVICE HAS BEEN RETURNED TO THE FACTORY AND IS BEING EVALUATED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED.
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIW 7XB CUTTING DEVICE DID NOT WORK. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIW 7XB CUTTING DEVICE DID NOT WORK. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 549487 | VASOVIEW 7 XB | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | MAQUET CV | 25132654 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |