FDA Adverse Event Malfunction Summary report: N

ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM

MDR report key: 6763619 · Received August 3, 2017

Report

Report Number
3006575795-2017-00251
Event Type
Malfunction
Date Received
August 3, 2017
Date of Event
June 15, 2017
Report Date
September 15, 2017
Manufacturer
ZYNO MEDICAL LLC
Product Code
FRN
UDI-DI
00814371020006
PMA / PMN Number
K130690
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE USER REPORTED OVER-INFUSING ISSUE WAS NOT CONFIRMED. THE DEVICE WAS FOUND TO HAVE A FLOW RATE ACCURACY OF -5.91%, WHICH WAS AN UNDER-INFUSING ISSUE AND WAS OUTSIDE OF THE +/-5% SPECIFICATION. THE DEVICE WAS ALSO FOUND TO HAVE A STIFF DOOR HINGE AND A NOISY MOTOR DURING INFUSION; NEITHER OF THESE ISSUES WAS RELATED TO THE REPORTED FLOW RATE ISSUE. THE PUMP WAS REPAIRED, RECALIBRATED, AND TESTED TO MEET ALL SPECIFICATION ON 09/06/2017.

Additional Manufacturer Narrative · 1

THE MANUFACTURER IS CURRENTLY WAITING FOR THE ARRIVAL OF THE DEVICE.

Description of Event or Problem · 1

THIS IS A FOLLOW-UP FOR THE INITIALLY FILED MDR (3006575795-2017-00251).

Description of Event or Problem · 1

THE USER REPORTED THAT THE DEVICE HAD AN OVER-INFUSING ISSUE. THE CUSTOMER REPORTED "ZYNO Z-800 - SN: (B)(4) ORIGINAL CAME IN FOR ACCURACY, VOLUME INFUSES FASTER THAN PROGRAMMED TIME (BAG PROGRAMMED TO INFUSE OVER 1 HR AND IS EMPTY AFTER 30 MINUTES BUT A REMAINING VOLUME STILL REGISTERED ON THE PUMP. WE DID FIND THAT THE PUMP HAS ACCURACY ISSUES, THE DOOR HINGE IS BENT MAKING OPENING AND CLOSING THE DOOR DIFFICULT. WE ARE UNABLE TO REPAIR THIS PUMP DUE TO THE 12 TOOTH STRUCTURE OF THE PUMP'S MECHANISM. THIS PUMP NEEDS TO GO INTO THE MANUFACTURE FOR REPAIRS." NO PATIENT INJURY OR HARM OCCURED FOR THIS CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
544928 ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM LARGE VOLUME INFUSION PUMP FRN ZYNO MEDICAL LLC Z-800 20110516-SH 00814371020006

Patients

Seq Age Sex Outcome Treatment
1