FDA Adverse Event Malfunction Summary report: N

RELION SK

MDR report key: 6760137 · Received August 2, 2017

Report

Report Number
1000113657-2017-01503
Event Type
Malfunction
Date Received
August 2, 2017
Date of Event
July 10, 2017
Report Date
August 2, 2017
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K051147
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) (B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20-USER'S TEST STRIP HAD POOR STORAGE (KITCHEN).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH AND LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH ALL RESULTS OBTAINED: 164, 121, 154, 208 AND 91 MG/DL. THE CUSTOMER REPORTED A FASTING METER TO METER COMPARISON TEST THAT PRODUCED RESULT OF 121 MG/DL USING THE RELION SK METER AND A RESULT OF 164 MG/DL USING DR.'S DEVICE (CUSTOMER HAD ALSO RECEIVED A RESULT OF 164 MG/DL USING RELION SK METER THAT IS IN THE METER MEMORY). THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 110 - 115 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/21/2019 AND OPEN VIAL DATE IS (B)(6) 2017. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (CUSTOMER IS CONCERNED WITH ALL FIVE RESULTS IN THE MEMORY): (B)(6). MEMORY CONCERNS:CUSTOMER CONCERNED WITH ALL THE 5 RESULTS PROVIDED INT HEMETER'S MEMORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
540772 RELION SK BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. RELION SK SU5092

Patients

Seq Age Sex Outcome Treatment
1 0 YR