RELION SK
Report
- Report Number
- 1000113657-2017-01503
- Event Type
- Malfunction
- Date Received
- August 2, 2017
- Date of Event
- July 10, 2017
- Report Date
- August 2, 2017
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K051147
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) (B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20-USER'S TEST STRIP HAD POOR STORAGE (KITCHEN).
CONSUMER REPORTED COMPLAINT FOR HIGH AND LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH ALL RESULTS OBTAINED: 164, 121, 154, 208 AND 91 MG/DL. THE CUSTOMER REPORTED A FASTING METER TO METER COMPARISON TEST THAT PRODUCED RESULT OF 121 MG/DL USING THE RELION SK METER AND A RESULT OF 164 MG/DL USING DR.'S DEVICE (CUSTOMER HAD ALSO RECEIVED A RESULT OF 164 MG/DL USING RELION SK METER THAT IS IN THE METER MEMORY). THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 110 - 115 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/21/2019 AND OPEN VIAL DATE IS (B)(6) 2017. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (CUSTOMER IS CONCERNED WITH ALL FIVE RESULTS IN THE MEMORY): (B)(6). MEMORY CONCERNS:CUSTOMER CONCERNED WITH ALL THE 5 RESULTS PROVIDED INT HEMETER'S MEMORY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 540772 | RELION SK | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | RELION SK | SU5092 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |