CERTAS INLINE VLV W/ SPHNGRD
Report
- Report Number
- 1226348-2017-10533
- Event Type
- Malfunction
- Date Received
- August 1, 2017
- Date of Event
- July 14, 2017
- Manufacturer
- CODMAN & SHURTLEFF, INC. / MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- K112156
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE PREVIOUSLY FILED REPORT LISTED AN INCORRECT LOT AND PRODUCT CODE COMBINATION FOR THE DHR REVIEW. THE REPORT SHOULD HAVE STATED: "REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-8805 WITH LOT 129402, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK ON THE 7TH APRIL 2017." THIS REPORT HAS BEEN UPDATED TO REFLECT THE CORRECTED INFORMATION. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.
(B)(4). IT WAS NOT POSSIBLE TO INVESTIGATE THE COMPLAINT AS NO SAMPLE WAS RETURNED FOR EVALUATION. NUMEROUS ATTEMPTS WERE MADE TO RECOVER THE DEVICE, WITH NO SUCCESS. IF THE SAMPLE IS RETURNED IN THE FUTURE, THIS COMPLAINT WILL BE RE-OPENED AND EVALUATED. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-8805 WITH LOT 129402, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK ON THE 7TH APRIL 2017.
IT WAS PREVIOUSLY REPORTED THAT THE DEVICE WOULD NOT BE RETURNED FOR EVALUATION. THE DEVICE WAS SUBSEQUENTLY RETURNED. THIS REPORT HAS BEEN UPDATED TO REFLECT THE CORRECTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
UPON COMPLETION OF THE INVESTIGATION IT WAS NOTED THAT THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS AT SETTING 3. THE VALVE WAS VISUALLY INSPECTED; NO DEFECTS WERE NOTED. THE VALVE WAS HYDRATED. THE VALVE WAS TESTED FOR PROGRAMMING. THE VALVE PASSED THE TEST. THE VALVE WAS FLUSHED, THE VALVE PASSED THE TEST NO OCCLUSION WAS NOTED. THE VALVE WAS LEAK TESTED, NO LEAKS NOTED. THE VALVE WAS REFLUX TESTED. THE VALVE PASSED THE TEST. THE SIPHON GUARD WAS TESTED. THE VALVE PASSED THE TEST. THE VALVE WAS DRIED. THE SIPHON GUARD WAS REMOVED. THE VALVE WAS THEN PRESSURE TESTED, THE VALVE PASSED THE TEST. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-3114 WITH LOT CKLBMT, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK ON THE 5TH APRIL 2017. NO ROOT CAUSE COULD BE DETERMINED AS NO FUNCTIONAL PROBLEM WAS NOTED WITH THE VALVES. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.
(B)(4). UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.
SHUNT VALVE COULDN'T BE PRESSED EFFICIENTLY. THE PUMP DIDN'T RISE AFTER PRESSING. THERE WAS A LEAKAGE AT THE END OF THE VALVE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 537777 | CERTAS INLINE VLV W/ SPHNGRD | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC. / MEDOS S.A. | 129402 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |