FDA Adverse Event Injury Summary report: N

BIOMONITOR 2-AF

MDR report key: 6754317 · Received July 31, 2017

Report

Report Number
1028232-2017-02667
Event Type
Injury
Date Received
July 31, 2017
Date of Event
July 7, 2017
Report Date
July 28, 2017
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
MXD
PMA / PMN Number
K152995
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

AN EROSION WAS OBSERVED FOLLOWING THE IMPLANTATION OF THIS BIOTRONIK DEVICE. THE STERILIZATION PROCESS WAS INVESTIGATED. THE VALIDATED PROCESS ASSURES THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., ARE WITHIN ITS SPECIFIED RANGES FOR EACH DISTRIBUTED DEVICE. ADDITIONALLY AN ANALYSIS OF VALIDATED MICROBIOLOGICAL INDICATORS IS PERFORMED AFTER EVERY STERILIZATION PROCEDURE AS EVIDENCE OF SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. REVIEW OF THE BIOTRONIK COMPLAINT DATABASE DID NOT REVEAL ANY CHANGES REGARDING THE TREND FOR THIS TYPE OF INCIDENT. IN SUMMARY, THE EROSION WAS NOT DEVICE RELATED.

Description of Event or Problem · 1

THE PATIENT WITH THIS DEVICE TWIDDLED THE DEVICE AND IT STARTED COMING OUT OF THE POCKET. THE PATIENTS HUSBAND REMOVED THE DEVICE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
534934 BIOMONITOR 2-AF IMPLANTABLE CARDIAC MONITOR MXD BIOTRONIK SE & CO. KG 398493

Patients

Seq Age Sex Outcome Treatment
1 72 YR Hospitalization