BIOMONITOR 2-AF
Report
- Report Number
- 1028232-2017-02667
- Event Type
- Injury
- Date Received
- July 31, 2017
- Date of Event
- July 7, 2017
- Report Date
- July 28, 2017
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- MXD
- PMA / PMN Number
- K152995
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
AN EROSION WAS OBSERVED FOLLOWING THE IMPLANTATION OF THIS BIOTRONIK DEVICE. THE STERILIZATION PROCESS WAS INVESTIGATED. THE VALIDATED PROCESS ASSURES THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., ARE WITHIN ITS SPECIFIED RANGES FOR EACH DISTRIBUTED DEVICE. ADDITIONALLY AN ANALYSIS OF VALIDATED MICROBIOLOGICAL INDICATORS IS PERFORMED AFTER EVERY STERILIZATION PROCEDURE AS EVIDENCE OF SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. REVIEW OF THE BIOTRONIK COMPLAINT DATABASE DID NOT REVEAL ANY CHANGES REGARDING THE TREND FOR THIS TYPE OF INCIDENT. IN SUMMARY, THE EROSION WAS NOT DEVICE RELATED.
THE PATIENT WITH THIS DEVICE TWIDDLED THE DEVICE AND IT STARTED COMING OUT OF THE POCKET. THE PATIENTS HUSBAND REMOVED THE DEVICE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 534934 | BIOMONITOR 2-AF | IMPLANTABLE CARDIAC MONITOR | MXD | BIOTRONIK SE & CO. KG | 398493 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization |