FDA Adverse Event
Death
Summary report: N
FILTA THERM
MDR report key: 67540
·
Received February 6, 1997
Report
- Report Number
- 67540
- Event Type
- Death
- Date Received
- February 6, 1997
- Date of Event
- February 2, 1997
- Report Date
- February 6, 1997
- Manufacturer
- INTERSURGICAL, INC.
- Product Code
- BTT
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- WI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
UPON ARRIVAL FOR ROUTINE VENTILATOR CHECK AND TREATMENT, FOUND PT DISCONNECTED FROM THE VENT. NO AUDIBLE OR VISUAL ALARMS. VENTILATOR FUNCTIONAL CHECK AFTER REMOVAL FROM PT. LOW PRESSURE ALARM SET TO 12CM H2O. WITH VENTILATOR CYCLING, RESISTANCE IN THE CIRCUIT EXCEEDED 10-12CM H2O. WHEN HEAT MOISTURE EXCHANGE FILTER REMOVED FROM THE CIRCUIT, THE LOW PRESSURE ALARM SOUNDED. CONCLUDE THAT RESISTANCE IN HME EXCEEDED THE LOW PRESSURE ALARM SETTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FILTA THERM | HUMIDIFIERS, HEAT/MOISTURE EXCHANGE | BTT | INTERSURGICAL, INC. | FILTA-THERM | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Death |