FDA Adverse Event Death Summary report: N

FILTA THERM

MDR report key: 67540 · Received February 6, 1997

Report

Report Number
67540
Event Type
Death
Date Received
February 6, 1997
Date of Event
February 2, 1997
Report Date
February 6, 1997
Manufacturer
INTERSURGICAL, INC.
Product Code
BTT
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
WI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

UPON ARRIVAL FOR ROUTINE VENTILATOR CHECK AND TREATMENT, FOUND PT DISCONNECTED FROM THE VENT. NO AUDIBLE OR VISUAL ALARMS. VENTILATOR FUNCTIONAL CHECK AFTER REMOVAL FROM PT. LOW PRESSURE ALARM SET TO 12CM H2O. WITH VENTILATOR CYCLING, RESISTANCE IN THE CIRCUIT EXCEEDED 10-12CM H2O. WHEN HEAT MOISTURE EXCHANGE FILTER REMOVED FROM THE CIRCUIT, THE LOW PRESSURE ALARM SOUNDED. CONCLUDE THAT RESISTANCE IN HME EXCEEDED THE LOW PRESSURE ALARM SETTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FILTA THERM HUMIDIFIERS, HEAT/MOISTURE EXCHANGE BTT INTERSURGICAL, INC. FILTA-THERM *

Patients

Seq Age Sex Outcome Treatment
1 75 YR Death